AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 0x0613
Error Code System Message
Question
What does 0x0613 mean on a GE Brivo XR285amx?
Short Answer
0x0613: RTL 4 error.
Possible Causes
The Djinn Beetle generator reported a kV FW fault: RTL 4 error.
Troubleshooting Steps
Run KV Measurement, KV Test, Inverter Test, and Resonant Components as applicable. Check the HV Tank-to-Power Module measurement cable, DC-bus error data, inverter/tank connections, and HV cables for arcing when indicated.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0614
Error Code System Message
Question
What does 0x0614 mean on a GE Brivo XR285amx?
Short Answer
0x0614: Watchdog error - Rose RT task is stopped.
Possible Causes
The Djinn Beetle generator reported a kV FW fault: Watchdog error - Rose RT task is stopped.
Troubleshooting Steps
Run KV Measurement, KV Test, Inverter Test, and Resonant Components as applicable. Check the HV Tank-to-Power Module measurement cable, DC-bus error data, inverter/tank connections, and HV cables for arcing when indicated.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0615
Error Code System Message
Question
What does 0x0615 mean on a GE Brivo XR285amx?
Short Answer
0x0615: ADC frozen.
Possible Causes
The Djinn Beetle generator reported a kV FW fault: ADC frozen.
Troubleshooting Steps
Run KV Measurement, KV Test, Inverter Test, and Resonant Components as applicable. Check the HV Tank-to-Power Module measurement cable, DC-bus error data, inverter/tank connections, and HV cables for arcing when indicated.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0616
Error Code System Message
Question
What does 0x0616 mean on a GE Brivo XR285amx?
Short Answer
0x0616: Resonant limit reached too many times.
Possible Causes
The Djinn Beetle generator reported a kV FW fault: Resonant limit reached too many times.
Troubleshooting Steps
Run KV Measurement, KV Test, Inverter Test, and Resonant Components as applicable. Check the HV Tank-to-Power Module measurement cable, DC-bus error data, inverter/tank connections, and HV cables for arcing when indicated.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0681
Error Code System Message
Question
What does 0x0681 mean on a GE Brivo XR285amx?
Short Answer
0x0681: Exp cmd while not ready.
Possible Causes
The Djinn Beetle generator reported a kV SW fault: Exp cmd while not ready.
Troubleshooting Steps
Run KV Measurement, KV Test, Inverter Test, and Resonant Components as applicable. Check the HV Tank-to-Power Module measurement cable, DC-bus error data, inverter/tank connections, and HV cables for arcing when indicated.
Parts Areas to Check