top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 0x0613

Error Code System Message

​

0x0613, 0X0613, Brivo XR285amx 0x0613, Djinn kV error, RTL 4 error
Question

What does 0x0613 mean on a GE Brivo XR285amx?

Short Answer

0x0613: RTL 4 error.

Possible Causes

The Djinn Beetle generator reported a kV FW fault: RTL 4 error.

Troubleshooting Steps

Run KV Measurement, KV Test, Inverter Test, and Resonant Components as applicable. Check the HV Tank-to-Power Module measurement cable, DC-bus error data, inverter/tank connections, and HV cables for arcing when indicated.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0614

Error Code System Message

​

0x0614, 0X0614, Brivo XR285amx 0x0614, Djinn kV error, Watchdog error - Rose RT task is stopped
Question

What does 0x0614 mean on a GE Brivo XR285amx?

Short Answer

0x0614: Watchdog error - Rose RT task is stopped.

Possible Causes

The Djinn Beetle generator reported a kV FW fault: Watchdog error - Rose RT task is stopped.

Troubleshooting Steps

Run KV Measurement, KV Test, Inverter Test, and Resonant Components as applicable. Check the HV Tank-to-Power Module measurement cable, DC-bus error data, inverter/tank connections, and HV cables for arcing when indicated.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0615

Error Code System Message

​

0x0615, 0X0615, Brivo XR285amx 0x0615, Djinn kV error, ADC frozen
Question

What does 0x0615 mean on a GE Brivo XR285amx?

Short Answer

0x0615: ADC frozen.

Possible Causes

The Djinn Beetle generator reported a kV FW fault: ADC frozen.

Troubleshooting Steps

Run KV Measurement, KV Test, Inverter Test, and Resonant Components as applicable. Check the HV Tank-to-Power Module measurement cable, DC-bus error data, inverter/tank connections, and HV cables for arcing when indicated.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0616

Error Code System Message

​

0x0616, 0X0616, Brivo XR285amx 0x0616, Djinn kV error, Resonant limit reached too many times
Question

What does 0x0616 mean on a GE Brivo XR285amx?

Short Answer

0x0616: Resonant limit reached too many times.

Possible Causes

The Djinn Beetle generator reported a kV FW fault: Resonant limit reached too many times.

Troubleshooting Steps

Run KV Measurement, KV Test, Inverter Test, and Resonant Components as applicable. Check the HV Tank-to-Power Module measurement cable, DC-bus error data, inverter/tank connections, and HV cables for arcing when indicated.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0681

Error Code System Message

​

0x0681, 0X0681, Brivo XR285amx 0x0681, Djinn kV error, Exp cmd while not ready
Question

What does 0x0681 mean on a GE Brivo XR285amx?

Short Answer

0x0681: Exp cmd while not ready.

Possible Causes

The Djinn Beetle generator reported a kV SW fault: Exp cmd while not ready.

Troubleshooting Steps

Run KV Measurement, KV Test, Inverter Test, and Resonant Components as applicable. Check the HV Tank-to-Power Module measurement cable, DC-bus error data, inverter/tank connections, and HV cables for arcing when indicated.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

bottom of page