top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 0x0682

Error Code System Message

​

0x0682, 0X0682, Brivo XR285amx 0x0682, Djinn kV error, Power module internal cable missing or damaged
Question

What does 0x0682 mean on a GE Brivo XR285amx?

Short Answer

0x0682: Power module internal cable missing or damaged.

Possible Causes

The Djinn Beetle generator reported a kV SW fault: Power module internal cable missing or damaged.

Troubleshooting Steps

Run KV Measurement, KV Test, Inverter Test, and Resonant Components as applicable. Check the HV Tank-to-Power Module measurement cable, DC-bus error data, inverter/tank connections, and HV cables for arcing when indicated.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0683

Error Code System Message

​

0x0683, 0X0683, Brivo XR285amx 0x0683, Djinn kV error, Measurement cable between tank and power missing or damaged error
Question

What does 0x0683 mean on a GE Brivo XR285amx?

Short Answer

0x0683: Measurement cable between tank and power missing or damaged error.

Possible Causes

The Djinn Beetle generator reported a kV SW fault: Measurement cable between tank and power missing or damaged error.

Troubleshooting Steps

Run KV Measurement, KV Test, Inverter Test, and Resonant Components as applicable. Check the HV Tank-to-Power Module measurement cable, DC-bus error data, inverter/tank connections, and HV cables for arcing when indicated.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0684

Error Code System Message

​

0x0684, 0X0684, Brivo XR285amx 0x0684, Djinn kV error, kV detected while inverter is OFF
Question

What does 0x0684 mean on a GE Brivo XR285amx?

Short Answer

0x0684: kV detected while inverter is OFF.

Possible Causes

The Djinn Beetle generator reported a kV SW fault: kV detected while inverter is OFF.

Troubleshooting Steps

Run KV Measurement, KV Test, Inverter Test, and Resonant Components as applicable. Check the HV Tank-to-Power Module measurement cable, DC-bus error data, inverter/tank connections, and HV cables for arcing when indicated.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0685

Error Code System Message

​

0x0685, 0X0685, Brivo XR285amx 0x0685, Djinn kV error, ILR not zero while OFF
Question

What does 0x0685 mean on a GE Brivo XR285amx?

Short Answer

0x0685: ILR not zero while OFF.

Possible Causes

The Djinn Beetle generator reported a kV SW fault: ILR not zero while OFF.

Troubleshooting Steps

Run KV Measurement, KV Test, Inverter Test, and Resonant Components as applicable. Check the HV Tank-to-Power Module measurement cable, DC-bus error data, inverter/tank connections, and HV cables for arcing when indicated.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0686

Error Code System Message

​

0x0686, 0X0686, Brivo XR285amx 0x0686, Djinn kV error, Measure connection failure
Question

What does 0x0686 mean on a GE Brivo XR285amx?

Short Answer

0x0686: Measure connection failure.

Possible Causes

The Djinn Beetle generator reported a kV SW fault: Measure connection failure.

Troubleshooting Steps

Run KV Measurement, KV Test, Inverter Test, and Resonant Components as applicable. Check the HV Tank-to-Power Module measurement cable, DC-bus error data, inverter/tank connections, and HV cables for arcing when indicated.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

bottom of page