AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 0x0701
Error Code System Message
Question
What does 0x0701 mean on a GE Brivo XR285amx?
Short Answer
0x0701: Tube overtemp switch open.
Possible Causes
The Djinn Beetle generator reported a Thermal SW fault: Tube overtemp switch open.
Troubleshooting Steps
Run Power Supply Thermal and the applicable thermal sensor tests. Check cooling, fans, sensor wiring, tube thermal-switch resistance, and single-wire communication before replacing a module.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0702
Error Code System Message
Question
What does 0x0702 mean on a GE Brivo XR285amx?
Short Answer
0x0702: Tank temperature too high.
Possible Causes
The Djinn Beetle generator reported a Thermal SW fault: Tank temperature too high.
Troubleshooting Steps
Run Power Supply Thermal and the applicable thermal sensor tests. Check cooling, fans, sensor wiring, tube thermal-switch resistance, and single-wire communication before replacing a module.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0703
Error Code System Message
Question
What does 0x0703 mean on a GE Brivo XR285amx?
Short Answer
0x0703: Inverter temperature too high.
Possible Causes
The Djinn Beetle generator reported a Thermal SW fault: Inverter temperature too high.
Troubleshooting Steps
Run Power Supply Thermal and the applicable thermal sensor tests. Check cooling, fans, sensor wiring, tube thermal-switch resistance, and single-wire communication before replacing a module.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0704
Error Code System Message
Question
What does 0x0704 mean on a GE Brivo XR285amx?
Short Answer
0x0704: Filament-Bias temperature too high.
Possible Causes
The Djinn Beetle generator reported a Thermal SW fault: Filament-Bias temperature too high.
Troubleshooting Steps
Run Power Supply Thermal and the applicable thermal sensor tests. Check cooling, fans, sensor wiring, tube thermal-switch resistance, and single-wire communication before replacing a module.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0705
Error Code System Message
Question
What does 0x0705 mean on a GE Brivo XR285amx?
Short Answer
0x0705: Rotor board temperature too high.
Possible Causes
The Djinn Beetle generator reported a Thermal SW fault: Rotor board temperature too high.
Troubleshooting Steps
Run Power Supply Thermal and the applicable thermal sensor tests. Check cooling, fans, sensor wiring, tube thermal-switch resistance, and single-wire communication before replacing a module.
Parts Areas to Check