top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 0x0782

Error Code System Message

​

0x0782, 0X0782, Brivo XR285amx 0x0782, Djinn Thermal error, Tank sensor failure
Question

What does 0x0782 mean on a GE Brivo XR285amx?

Short Answer

0x0782: Tank sensor failure.

Possible Causes

The Djinn Beetle generator reported a Thermal SW fault: Tank sensor failure.

Troubleshooting Steps

Run Power Supply Thermal and the applicable thermal sensor tests. Check cooling, fans, sensor wiring, tube thermal-switch resistance, and single-wire communication before replacing a module.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0783

Error Code System Message

​

0x0783, 0X0783, Brivo XR285amx 0x0783, Djinn Thermal error, Inverter sensor failure
Question

What does 0x0783 mean on a GE Brivo XR285amx?

Short Answer

0x0783: Inverter sensor failure.

Possible Causes

The Djinn Beetle generator reported a Thermal SW fault: Inverter sensor failure.

Troubleshooting Steps

Run Power Supply Thermal and the applicable thermal sensor tests. Check cooling, fans, sensor wiring, tube thermal-switch resistance, and single-wire communication before replacing a module.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0784

Error Code System Message

​

0x0784, 0X0784, Brivo XR285amx 0x0784, Djinn Thermal error, Tube Thermal protection
Question

What does 0x0784 mean on a GE Brivo XR285amx?

Short Answer

0x0784: Tube Thermal protection.

Possible Causes

The Djinn Beetle generator reported a Thermal SW fault: Tube Thermal protection.

Troubleshooting Steps

Run Power Supply Thermal and the applicable thermal sensor tests. Check cooling, fans, sensor wiring, tube thermal-switch resistance, and single-wire communication before replacing a module.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0785

Error Code System Message

​

0x0785, 0X0785, Brivo XR285amx 0x0785, Djinn Thermal error, Single wire device configuration attempt failure
Question

What does 0x0785 mean on a GE Brivo XR285amx?

Short Answer

0x0785: Single wire device configuration attempt failure.

Possible Causes

The Djinn Beetle generator reported a Thermal SW fault: Single wire device configuration attempt failure.

Troubleshooting Steps

Run Power Supply Thermal and the applicable thermal sensor tests. Check cooling, fans, sensor wiring, tube thermal-switch resistance, and single-wire communication before replacing a module.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0786

Error Code System Message

​

0x0786, 0X0786, Brivo XR285amx 0x0786, Djinn Thermal error, Single Wire device not correctly configured
Question

What does 0x0786 mean on a GE Brivo XR285amx?

Short Answer

0x0786: Single Wire device not correctly configured.

Possible Causes

The Djinn Beetle generator reported a Thermal SW fault: Single Wire device not correctly configured.

Troubleshooting Steps

Run Power Supply Thermal and the applicable thermal sensor tests. Check cooling, fans, sensor wiring, tube thermal-switch resistance, and single-wire communication before replacing a module.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

bottom of page