AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 0x0782
Error Code System Message
Question
What does 0x0782 mean on a GE Brivo XR285amx?
Short Answer
0x0782: Tank sensor failure.
Possible Causes
The Djinn Beetle generator reported a Thermal SW fault: Tank sensor failure.
Troubleshooting Steps
Run Power Supply Thermal and the applicable thermal sensor tests. Check cooling, fans, sensor wiring, tube thermal-switch resistance, and single-wire communication before replacing a module.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0783
Error Code System Message
Question
What does 0x0783 mean on a GE Brivo XR285amx?
Short Answer
0x0783: Inverter sensor failure.
Possible Causes
The Djinn Beetle generator reported a Thermal SW fault: Inverter sensor failure.
Troubleshooting Steps
Run Power Supply Thermal and the applicable thermal sensor tests. Check cooling, fans, sensor wiring, tube thermal-switch resistance, and single-wire communication before replacing a module.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0784
Error Code System Message
Question
What does 0x0784 mean on a GE Brivo XR285amx?
Short Answer
0x0784: Tube Thermal protection.
Possible Causes
The Djinn Beetle generator reported a Thermal SW fault: Tube Thermal protection.
Troubleshooting Steps
Run Power Supply Thermal and the applicable thermal sensor tests. Check cooling, fans, sensor wiring, tube thermal-switch resistance, and single-wire communication before replacing a module.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0785
Error Code System Message
Question
What does 0x0785 mean on a GE Brivo XR285amx?
Short Answer
0x0785: Single wire device configuration attempt failure.
Possible Causes
The Djinn Beetle generator reported a Thermal SW fault: Single wire device configuration attempt failure.
Troubleshooting Steps
Run Power Supply Thermal and the applicable thermal sensor tests. Check cooling, fans, sensor wiring, tube thermal-switch resistance, and single-wire communication before replacing a module.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0786
Error Code System Message
Question
What does 0x0786 mean on a GE Brivo XR285amx?
Short Answer
0x0786: Single Wire device not correctly configured.
Possible Causes
The Djinn Beetle generator reported a Thermal SW fault: Single Wire device not correctly configured.
Troubleshooting Steps
Run Power Supply Thermal and the applicable thermal sensor tests. Check cooling, fans, sensor wiring, tube thermal-switch resistance, and single-wire communication before replacing a module.
Parts Areas to Check