top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 0x0A90

Error Code System Message

​

0x0A90, 0X0A90, Brivo XR285amx 0x0A90, Djinn Various error, NVRAM battery is low
Question

What does 0x0A90 mean on a GE Brivo XR285amx?

Short Answer

0x0A90: NVRAM battery is low.

Possible Causes

The Djinn Beetle generator reported a Various SW fault: NVRAM battery is low.

Troubleshooting Steps

Run System Sanity and Power Supplies, then review the error log. Check the indicated internal power, NVRAM/RTC, FPGA communication, or reboot condition before isolating an FRU.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0A91

Error Code System Message

​

0x0A91, 0X0A91, Brivo XR285amx 0x0A91, Djinn Various error, Real-time clock not working
Question

What does 0x0A91 mean on a GE Brivo XR285amx?

Short Answer

0x0A91: Real-time clock not working.

Possible Causes

The Djinn Beetle generator reported a Various SW fault: Real-time clock not working.

Troubleshooting Steps

Run System Sanity and Power Supplies, then review the error log. Check the indicated internal power, NVRAM/RTC, FPGA communication, or reboot condition before isolating an FRU.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0AA0

Error Code System Message

​

0x0AA0, 0X0AA0, Brivo XR285amx 0x0AA0, Djinn Various error, Reboot caused by an unexpected exception
Question

What does 0x0AA0 mean on a GE Brivo XR285amx?

Short Answer

0x0AA0: Reboot caused by an unexpected exception.

Possible Causes

The Djinn Beetle generator reported a Various SW fault: Reboot caused by an unexpected exception.

Troubleshooting Steps

Run System Sanity and Power Supplies, then review the error log. Check the indicated internal power, NVRAM/RTC, FPGA communication, or reboot condition before isolating an FRU.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0AA1

Error Code System Message

​

0x0AA1, 0X0AA1, Brivo XR285amx 0x0AA1, Djinn Various error, FPGA communication watchdog error
Question

What does 0x0AA1 mean on a GE Brivo XR285amx?

Short Answer

0x0AA1: FPGA communication watchdog error.

Possible Causes

The Djinn Beetle generator reported a Various SW fault: FPGA communication watchdog error.

Troubleshooting Steps

Run System Sanity and Power Supplies, then review the error log. Check the indicated internal power, NVRAM/RTC, FPGA communication, or reboot condition before isolating an FRU.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0B81

Error Code System Message

​

0x0B81, 0X0B81, Brivo XR285amx 0x0B81, Djinn mA error, mA measured while inverter is off
Question

What does 0x0B81 mean on a GE Brivo XR285amx?

Short Answer

0x0B81: mA measured while inverter is off.

Possible Causes

The Djinn Beetle generator reported a mA SW fault: mA measured while inverter is off.

Troubleshooting Steps

Run mA Measurement and mA Test. Reseat the HV Tank J6-to-Power Module J7 measurement cable and perform the specified continuity/resistance checks; use filament ageing clearing/calibration when 0x0B84-0x0B86 is involved.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

bottom of page