AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 0x0A90
Error Code System Message
Question
What does 0x0A90 mean on a GE Brivo XR285amx?
Short Answer
0x0A90: NVRAM battery is low.
Possible Causes
The Djinn Beetle generator reported a Various SW fault: NVRAM battery is low.
Troubleshooting Steps
Run System Sanity and Power Supplies, then review the error log. Check the indicated internal power, NVRAM/RTC, FPGA communication, or reboot condition before isolating an FRU.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0A91
Error Code System Message
Question
What does 0x0A91 mean on a GE Brivo XR285amx?
Short Answer
0x0A91: Real-time clock not working.
Possible Causes
The Djinn Beetle generator reported a Various SW fault: Real-time clock not working.
Troubleshooting Steps
Run System Sanity and Power Supplies, then review the error log. Check the indicated internal power, NVRAM/RTC, FPGA communication, or reboot condition before isolating an FRU.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0AA0
Error Code System Message
Question
What does 0x0AA0 mean on a GE Brivo XR285amx?
Short Answer
0x0AA0: Reboot caused by an unexpected exception.
Possible Causes
The Djinn Beetle generator reported a Various SW fault: Reboot caused by an unexpected exception.
Troubleshooting Steps
Run System Sanity and Power Supplies, then review the error log. Check the indicated internal power, NVRAM/RTC, FPGA communication, or reboot condition before isolating an FRU.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0AA1
Error Code System Message
Question
What does 0x0AA1 mean on a GE Brivo XR285amx?
Short Answer
0x0AA1: FPGA communication watchdog error.
Possible Causes
The Djinn Beetle generator reported a Various SW fault: FPGA communication watchdog error.
Troubleshooting Steps
Run System Sanity and Power Supplies, then review the error log. Check the indicated internal power, NVRAM/RTC, FPGA communication, or reboot condition before isolating an FRU.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0B81
Error Code System Message
Question
What does 0x0B81 mean on a GE Brivo XR285amx?
Short Answer
0x0B81: mA measured while inverter is off.
Possible Causes
The Djinn Beetle generator reported a mA SW fault: mA measured while inverter is off.
Troubleshooting Steps
Run mA Measurement and mA Test. Reseat the HV Tank J6-to-Power Module J7 measurement cable and perform the specified continuity/resistance checks; use filament ageing clearing/calibration when 0x0B84-0x0B86 is involved.
Parts Areas to Check