Error Number

Error 0x0B82

Error Code System Message

​

0x0B82, 0X0B82, Brivo XR285amx 0x0B82, Djinn mA error, No mA measured at the beginning of the pulse
Question

What does 0x0B82 mean on a GE Brivo XR285amx?

Short Answer

0x0B82: No mA measured at the beginning of the pulse.

Possible Causes

The Djinn Beetle generator reported a mA SW fault: No mA measured at the beginning of the pulse.

Troubleshooting Steps

Run mA Measurement and mA Test. Reseat the HV Tank J6-to-Power Module J7 measurement cable and perform the specified continuity/resistance checks; use filament ageing clearing/calibration when 0x0B84-0x0B86 is involved.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0B83

Error Code System Message

​

0x0B83, 0X0B83, Brivo XR285amx 0x0B83, Djinn mA error, Inconsistency between mA anode and mA cathode
Question

What does 0x0B83 mean on a GE Brivo XR285amx?

Short Answer

0x0B83: Inconsistency between mA anode and mA cathode.

Possible Causes

The Djinn Beetle generator reported a mA SW fault: Inconsistency between mA anode and mA cathode.

Troubleshooting Steps

Run mA Measurement and mA Test. Reseat the HV Tank J6-to-Power Module J7 measurement cable and perform the specified continuity/resistance checks; use filament ageing clearing/calibration when 0x0B84-0x0B86 is involved.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0B84

Error Code System Message

​

0x0B84, 0X0B84, Brivo XR285amx 0x0B84, Djinn mA error, mA too low
Question

What does 0x0B84 mean on a GE Brivo XR285amx?

Short Answer

0x0B84: mA too low.

Possible Causes

The Djinn Beetle generator reported a mA SW fault: mA too low.

Troubleshooting Steps

Run mA Measurement and mA Test. Reseat the HV Tank J6-to-Power Module J7 measurement cable and perform the specified continuity/resistance checks; use filament ageing clearing/calibration when 0x0B84-0x0B86 is involved.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0B85

Error Code System Message

​

0x0B85, 0X0B85, Brivo XR285amx 0x0B85, Djinn mA error, mA too high
Question

What does 0x0B85 mean on a GE Brivo XR285amx?

Short Answer

0x0B85: mA too high.

Possible Causes

The Djinn Beetle generator reported a mA SW fault: mA too high.

Troubleshooting Steps

Run mA Measurement and mA Test. Reseat the HV Tank J6-to-Power Module J7 measurement cable and perform the specified continuity/resistance checks; use filament ageing clearing/calibration when 0x0B84-0x0B86 is involved.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0B86

Error Code System Message

​

0x0B86, 0X0B86, Brivo XR285amx 0x0B86, Djinn mA error, Filament ageing saturated
Question

What does 0x0B86 mean on a GE Brivo XR285amx?

Short Answer

0x0B86: Filament ageing saturated.

Possible Causes

The Djinn Beetle generator reported a mA SW fault: Filament ageing saturated.

Troubleshooting Steps

Run mA Measurement and mA Test. Reseat the HV Tank J6-to-Power Module J7 measurement cable and perform the specified continuity/resistance checks; use filament ageing clearing/calibration when 0x0B84-0x0B86 is involved.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

bottom of page