Error Number
Error 0x0B82
Error Code System Message
Question
What does 0x0B82 mean on a GE Brivo XR285amx?
Short Answer
0x0B82: No mA measured at the beginning of the pulse.
Possible Causes
The Djinn Beetle generator reported a mA SW fault: No mA measured at the beginning of the pulse.
Troubleshooting Steps
Run mA Measurement and mA Test. Reseat the HV Tank J6-to-Power Module J7 measurement cable and perform the specified continuity/resistance checks; use filament ageing clearing/calibration when 0x0B84-0x0B86 is involved.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0B83
Error Code System Message
Question
What does 0x0B83 mean on a GE Brivo XR285amx?
Short Answer
0x0B83: Inconsistency between mA anode and mA cathode.
Possible Causes
The Djinn Beetle generator reported a mA SW fault: Inconsistency between mA anode and mA cathode.
Troubleshooting Steps
Run mA Measurement and mA Test. Reseat the HV Tank J6-to-Power Module J7 measurement cable and perform the specified continuity/resistance checks; use filament ageing clearing/calibration when 0x0B84-0x0B86 is involved.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0B84
Error Code System Message
Question
What does 0x0B84 mean on a GE Brivo XR285amx?
Short Answer
0x0B84: mA too low.
Possible Causes
The Djinn Beetle generator reported a mA SW fault: mA too low.
Troubleshooting Steps
Run mA Measurement and mA Test. Reseat the HV Tank J6-to-Power Module J7 measurement cable and perform the specified continuity/resistance checks; use filament ageing clearing/calibration when 0x0B84-0x0B86 is involved.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0B85
Error Code System Message
Question
What does 0x0B85 mean on a GE Brivo XR285amx?
Short Answer
0x0B85: mA too high.
Possible Causes
The Djinn Beetle generator reported a mA SW fault: mA too high.
Troubleshooting Steps
Run mA Measurement and mA Test. Reseat the HV Tank J6-to-Power Module J7 measurement cable and perform the specified continuity/resistance checks; use filament ageing clearing/calibration when 0x0B84-0x0B86 is involved.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0B86
Error Code System Message
Question
What does 0x0B86 mean on a GE Brivo XR285amx?
Short Answer
0x0B86: Filament ageing saturated.
Possible Causes
The Djinn Beetle generator reported a mA SW fault: Filament ageing saturated.
Troubleshooting Steps
Run mA Measurement and mA Test. Reseat the HV Tank J6-to-Power Module J7 measurement cable and perform the specified continuity/resistance checks; use filament ageing clearing/calibration when 0x0B84-0x0B86 is involved.
Parts Areas to Check