AMX Troubleshooting & Technical Support

Results Count

Error Number

Tube rotor issues

Error Code System Message

​

Brivo XR285 tube rotor impedance, stator resistance
Question

How do I test tube rotor impedance on the Brivo XR285amx?

Short Answer

The Tube Rotor Impedance diagnostic measures the tube stator winding impedance.

Possible Causes

Open/shorted stator winding; tube rotation cable issue; rotor-board fault.

Troubleshooting Steps

Run Tube Rotor Impedance. If abnormal, inspect the rotation cable, then measure stator Common-to-Phase and Common-to-Auxiliary resistance and isolation at the tube side.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Tube thermal-sensor fault

Error Code System Message

​

Brivo XR285 tube thermal switch, 0701 0784
Question

How do I isolate a Brivo XR285amx tube thermal-sensor fault?

Short Answer

Measure the tube thermal switch/cable resistance; the manual specifies less than 2 ohms and uses repeat measurement at the tube to isolate cable versus tube.

Possible Causes

Open/high-resistance thermal switch or feedback cable; tube thermal protection.

Troubleshooting Steps

Disconnect the specified thermal cable and measure resistance. If above 2 ohms, measure directly at the tube with the feedback cable removed to isolate the tube from the cable.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Workflow issues

Error Code System Message

​

Brivo XR285amx troubleshooting workflow, SUIF diagnostics, error log
Question

What is the recommended troubleshooting workflow for the Brivo XR285amx?

Short Answer

Start by reviewing the error log, then ask the operator when and how the problem occurs, and finally run targeted SUIF diagnostics for the affected subsystem.

Possible Causes

System fault may be electrical, communication, drive, generator, or software related.

Troubleshooting Steps

Review the error log. Interview the operator for failure conditions. Open SUIF Diagnostics and run the appropriate System Power, System Control, Drive and Brakes, or Generator test.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

kV measurement errors

Error Code System Message

​

Brivo XR285 kV measurement, 0683 0686
Question

How do I troubleshoot Brivo XR285amx kV measurement errors?

Short Answer

Use KV Measurement/KV Test and verify the HV Tank-to-Power Module measurement cable and specified resistor values.

Possible Causes

Measurement cable or resistor issue; Power Module/HV Tank fault; DC bus/inverter issue.

Troubleshooting Steps

Reseat the J6-J7 measurement cable, run KV Measurement/KV Test, and if needed perform the resistance/cable checks before isolating the tank or power module.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

mA measurement errors

Error Code System Message

​

Brivo XR285 mA measurement, 0B84 0B85 0B87
Question

How do I troubleshoot Brivo XR285amx mA measurement errors?

Short Answer

Use the mA Measurement diagnostic and verify the ribbon cable between HV Tank J6 and Power Module J7; the manual also gives resistance checks for the measurement network.

Possible Causes

Measurement cable fault; resistor network issue; HV tank or power-module measurement fault.

Troubleshooting Steps

Reseat the J6-J7 measurement cable and run mA Measurement. If it fails, perform continuity and resistance checks specified in the manual before replacing an FRU.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

bottom of page