AMX Troubleshooting & Technical Support
Results Count
Error Number
Detector LED status steady RED
Error Code System Message
Steady RED
Question
What does Steady RED mean on the Definium AMX 700 detector?
Short Answer
Detector error condition.
Possible Causes
Detector error condition
Troubleshooting Steps
Do not use the detector unless instructed by the OnLine Center; log the date/time and scenario.
Parts Areas to Check
TRAD digital detector
For Related Models
Definium AMX 700
Error Number
Detector LED status Green and Red out of phase
Error Code System Message
Green and Red out of phase
Question
What does Green and Red out of phase mean on the Definium AMX 700 detector?
Short Answer
No communication with host.
Possible Causes
No communication with host
Troubleshooting Steps
Troubleshoot detector power, host connection and Ethernet communication.
Parts Areas to Check
TRAD digital detector
For Related Models
Definium AMX 700
Error Number
Detector LED status Green/Red in phase - temperature warning
Error Code System Message
Green/Red in phase - temperature warning
Question
What does Green/Red in phase - temperature warning mean on the Definium AMX 700 detector?
Short Answer
Detector contact temperature warning.
Possible Causes
Detector contact temperature warning
Troubleshooting Steps
Allow cooling and assess detector temperature before continued use.
Parts Areas to Check
TRAD digital detector
For Related Models
Definium AMX 700
Error Number
Detector LED status Red ON with BIST
Error Code System Message
Red ON with BIST fail pattern
Question
What does Red ON with BIST fail pattern mean on the Definium AMX 700 detector?
Short Answer
Built In Self Test failure.
Possible Causes
Built In Self Test failure
Troubleshooting Steps
Do not continue normal use; investigate the detector and escalate.
Parts Areas to Check
TRAD digital detector
For Related Models
Definium AMX 700
Error Number
Detector LED status blinking GREEN
Error Code System Message
Blinking GREEN
Question
What does Blinking GREEN mean on the Definium AMX 700 detector?
Short Answer
Idle state.
Possible Causes
Idle state
Troubleshooting Steps
Detector is waiting for commands from the operator/system.
Parts Areas to Check
TRAD digital detector