top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Detector LED status blinking RED

Error Code System Message

Blinking RED

AMX 700 detector Blinking RED, TRAD detector LED, Detector warning
Question

What does Blinking RED mean on the Definium AMX 700 detector?

Short Answer

Detector warning.

Possible Causes

Detector warning

Troubleshooting Steps

Proceed with caution and log the date/time and scenario.

Parts Areas to Check

TRAD digital detector

For Related Models

Definium AMX 700

Error Number

Detector LED status both LEDs OFF

Error Code System Message

Both LEDs OFF

AMX 700 detector Both LEDs OFF, TRAD detector LED, Detector has no power
Question

What does Both LEDs OFF mean on the Definium AMX 700 detector?

Short Answer

Detector has no power.

Possible Causes

Detector has no power

Troubleshooting Steps

Close the exam and restart if the no-power state is unexpected.

Parts Areas to Check

TRAD digital detector

For Related Models

Definium AMX 700

Error Number

Detector LED status fast blinking GREEN

Error Code System Message

Fast blinking GREEN

AMX 700 detector Fast blinking GREEN, TRAD detector LED, Image transferring
Question

What does Fast blinking GREEN mean on the Definium AMX 700 detector?

Short Answer

Image transferring.

Possible Causes

Image transferring

Troubleshooting Steps

Allow transfer to complete; do not move the detector.

Parts Areas to Check

TRAD digital detector

For Related Models

Definium AMX 700

Error Number

Detector LED status steady GREEN

Error Code System Message

Steady GREEN

AMX 700 detector Steady GREEN, TRAD detector LED, Ready for acquisition or image transfer
Question

What does Steady GREEN mean on the Definium AMX 700 detector?

Short Answer

Ready for acquisition or image transfer.

Possible Causes

Ready for acquisition or image transfer

Troubleshooting Steps

Do not move the detector until the green LED begins blinking.

Parts Areas to Check

TRAD digital detector

For Related Models

Definium AMX 700

Error Number

Detector park status incorrect

Error Code System Message

DETPARK_SW

AMX 700 detector park switch 24V 0V
Question

How do I verify the detector-park switch on a Definium AMX 700?

Short Answer

Debug-header connection 12 should be 24 V with the detector parked and 0 V when it is not parked.

Possible Causes

Detector park switch fault; wiring issue; incorrect detector position.

Troubleshooting Steps

Measure DETPARK_SW at debug-header connection 12 while parking and unparking the detector. The voltage should change between 24 V and 0 V.

Parts Areas to Check

Detector park switch; Beetle Board wiring

For Related Models

Definium AMX 700

bottom of page