AMX Troubleshooting & Technical Support
Results Count
Error Number
TEMP - Over-temperature condition
Error Code System Message
TEMP - Over-temperature condition
Question
What does “TEMP” mean on a GE Definium AMX 700?
Short Answer
TEMP: Over-temperature condition.
Possible Causes
Subsystem/category: Beetle Board. Source classification: LED STATUS.
Troubleshooting Steps
Interpret the LED together with the system symptom and use the matching Beetle/power/exposure troubleshooting path.
Parts Areas to Check
Beetle Interface Board
For Related Models
Definium AMX 700
Error Number
TEMP LED
Error Code System Message
TEMP - DS2 RED
Question
What does the TEMP LED mean on the GE Definium AMX 700?
Short Answer
The Beetle Board detected an over-temperature condition.
Possible Causes
Over-temperature condition
Troubleshooting Steps
Use the LED state together with the system symptom. If the LED indicates an error, inhibit or temperature condition, perform the corresponding subsystem troubleshooting before replacing the board.
Parts Areas to Check
Beetle Interface Board
For Related Models
Definium AMX 700
Error Number
TRAD detector has no lights
Error Code System Message
Both detector LEDs OFF
Question
Why are there no LEDs on the Definium AMX 700 TRAD detector?
Short Answer
No detector LEDs are treated as a fault condition and can indicate loss of detector power or connection.
Possible Causes
Detector power unavailable; loose power connection; Ethernet/power connection issue.
Troubleshooting Steps
Check detector power and Ethernet connections. Review detector LED definitions and the detector power-supply status before replacing hardware.
Parts Areas to Check
TRAD digital detector; detector power supply; cables
For Related Models
Definium AMX 700
Error Number
X-RAY ON
Error Code System Message
X-RAY ON - X-ray ON signal active
Question
What does “X-RAY ON” mean on a GE Definium AMX 700?
Short Answer
X-RAY ON: X-ray ON signal active.
Possible Causes
Subsystem/category: Beetle Board. Source classification: LED STATUS.
Troubleshooting Steps
Interpret the LED together with the system symptom and use the matching Beetle/power/exposure troubleshooting path.
Parts Areas to Check
Beetle Interface Board
For Related Models
Definium AMX 700
Error Number
kV/mA cannot be adjusted from AMX or UIF
Error Code System Message
Question
Why can’t I adjust kV or mA on the Definium AMX 700?
Short Answer
The functional test expects kV and mA changes from both the AMX controls and the user interface to track correctly.
Possible Causes
PC protocol/control communication issue; Beetle Board or AMX control-path issue.
Troubleshooting Steps
Allow the PC to boot, enter an exam and protocol, then adjust kV/mA from the AMX and UIF. If only one control path fails, isolate that interface; if both fail, check PC/Beetle communication.
Parts Areas to Check
UIF; Beetle Board; AMX controller; PC