top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Degraded mode

Error Code System Message

​

XR220 degraded mode, XR200 black screen, Spyder PC video
Question

Why does an XR200amx or XR220amx enter Degraded Mode?

Short Answer

The Exclusive manual provides a flow for degraded-mode conditions involving PC power/boot, video communication, the Spyder board and related cables.

Possible Causes

PC power or boot failure; video cable fault; Spyder fault; software/firmware recovery issue.

Troubleshooting Steps

Check whether the PC power light is present and whether an external monitor has normal video. Verify PC supply voltage and reseat the video cable. Follow the degraded-mode flow before replacing PC or Spyder.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Detector dropped or impacted

Error Code System Message

​

XR220 dropped detector, FlashPad shock sensor
Question

How do I check an XR220amx detector after a drop or impact?

Short Answer

The detector diagnostics include Shock Sensors, which demonstrate detector sensor functionality using the defined service test method.

Possible Causes

Drop/impact; possible internal detector damage.

Troubleshooting Steps

Remove the detector from clinical use if damage is suspected. Inspect for physical damage and run the approved Shock Sensor and quality-assurance checks before returning it to service.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Detector lost communication

Error Code System Message

​

XR220 lost communication, FlashPad disconnects, detector out of range
Question

Why does the XR220amx show a Lost Communication message for the detector?

Short Answer

A Lost Communication message can occur when the detector is too far from the host radio or RF blockage/interference degrades the wireless link.

Possible Causes

Excessive distance; RF interference/blockage; wireless communication-path fault.

Troubleshooting Steps

Move the detector into normal operating distance, observe detector and host-radio LEDs, run Wireless Comm/Signal Strength and compare operation using tether or GbEthernet.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Detector temperature issues

Error Code System Message

​

XR220 detector temperature, FlashPad overheat diagnostic
Question

How do I check detector temperature on XR220amx?

Short Answer

The detector diagnostic set reports internal sensor temperatures against limits, and the Exclusive manual includes dedicated temperature/overheat diagnostics.

Possible Causes

Detector overheating; internal sensor fault; environmental temperature condition.

Troubleshooting Steps

Run the detector Temperature diagnostic, confirm proper environment/airflow and compare results across operating modes. Escalate persistent over-limit results.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 0100

Error Code System Message

​

0100, error 0100, XR200 0100, XR220 0100, Optima XR200amx error 0100, Optima XR220amx error 0100, Fil Small Unexpected generator firmware error
Question

What does error 0100 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0100: Fil Small Unexpected generator firmware error.

Possible Causes

Subsystem/category: Filament Small. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Review related generator errors. Run mA Measurement and the applicable filament-driver tests; inspect tube/filament connections and use the specific error workflow before replacing parts.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

bottom of page