AMX Troubleshooting & Technical Support
Results Count
Error Number
Drive goes the wrong direction or feedback is wrong
Error Code System Message
ERROR 214-217
Question
What causes AMX-Series drive-feedback errors?
Short Answer
Drive-feedback errors occur when feedback is opposite to the command, exceeds the command, or does not follow the command within expected tolerance.
Possible Causes
Drive feedback wiring; Hall-sensor/handle issue; drive-controller output or feedback circuitry.
Troubleshooting Steps
Compare command and feedback in the run-time diagnostic display. Verify handle Hall-sensor voltages and left/right drive-feedback signals, then inspect drive-controller connections and fuses.
Parts Areas to Check
Drive control board; handle transducers; feedback circuitry
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
During “CAL CHARGER” - Unit Works Up To 1.2 Volts - When Value Is Entered, Charger Shuts Off And External Voltmeter Goes To Zero, Then You Get An “X-RAY BYTE LIMIT”.
Error Code System Message
Question
How do I troubleshoot this AMX-4 Series issue: During “CAL CHARGER” - Unit Works Up To 1.2 Volts - When Value Is Entered, Charger Shuts Off And External Voltmeter Goes To Zero, Then You Get An “X-RAY BYTE LIMIT”?
Short Answer
Defective CPU Board
Possible Causes
Problem: During “CAL CHARGER” - Unit Works Up To 1.2 Volts - When Value Is Entered, Charger Shuts Off And External Voltmeter Goes To Zero, Then You Get An “X-RAY BYTE LIMIT”.
Troubleshooting Steps
Defective CPU Board
Parts Areas to Check
CPU Board
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
Error 101
Error Code System Message
101
Question
What should I check for errors 101 on the AMX-4 Series?
Short Answer
1) Perform recalibration of the “CAL BATT CHARGER” portion. 2) Replace the Battery Charger Board.
Possible Causes
1) Perform recalibration of the “CAL BATT CHARGER” portion. 2) Replace the Battery Charger Board.
Troubleshooting Steps
1) Perform recalibration of the “CAL BATT CHARGER” portion. 2) Replace the Battery Charger Board.
Parts Areas to Check
Battery Charger Board; Battery
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
Error 101
Error Code System Message
ERROR 101 - Charge-current feedback high during high-current charging
Question
What does ERROR 101 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 101: Charge-current feedback high during high-current charging.
Possible Causes
Feedback exceeded expected high-current range.
Troubleshooting Steps
Check charger current-feedback/scale-select signals, CPU-to-charger connection and actual charge current.
Parts Areas to Check
Battery set; charger board; CPU board; voltage/current feedback wiring
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
Error 102
Error Code System Message
ERROR 102 - Charge-current feedback low during high-current charging
Question
What does ERROR 102 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 102: Charge-current feedback low during high-current charging.
Possible Causes
Feedback was below expected high-current range.
Troubleshooting Steps
Check charger feedback, scale-select signal, charger-board connection and charge current.
Parts Areas to Check
Battery set; charger board; CPU board; voltage/current feedback wiring