AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 0504
Error Code System Message
Question
What does error 0504 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0504: Input stage: DC Bus rising time too slow during DC Bus charge [DC Bus < 200V].
Possible Causes
Subsystem/category: Input Stage / DC Bus. Manual recovery/error class: ABORT.
Troubleshooting Steps
Check Power Control Limits and generator input/DC-bus diagnostics. Evaluate system battery state, input cabling and generator input-stage hardware.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 0505
Error Code System Message
Question
What does error 0505 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0505: Input stage: No input current detected during DC Bus charge.
Possible Causes
Subsystem/category: Input Stage / DC Bus. Manual recovery/error class: ABORT.
Troubleshooting Steps
Check Power Control Limits and generator input/DC-bus diagnostics. Evaluate system battery state, input cabling and generator input-stage hardware.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 0510
Error Code System Message
Question
What does error 0510 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0510: Input stage: Over voltage on DC output detected [DC Bus > 460V].
Possible Causes
Subsystem/category: Input Stage / DC Bus. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Check Power Control Limits and generator input/DC-bus diagnostics. Evaluate system battery state, input cabling and generator input-stage hardware.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 0511
Error Code System Message
Question
What does error 0511 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0511: Input stage: Unexpected generator firmware error.
Possible Causes
Subsystem/category: Input Stage / DC Bus. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Check Power Control Limits and generator input/DC-bus diagnostics. Evaluate system battery state, input cabling and generator input-stage hardware.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 0520
Error Code System Message
Question
What does error 0520 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0520: Input stage: Software over input current detected [Iin > 140A].
Possible Causes
Subsystem/category: Input Stage / DC Bus. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Check Power Control Limits and generator input/DC-bus diagnostics. Evaluate system battery state, input cabling and generator input-stage hardware.
Parts Areas to Check