top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 0504

Error Code System Message

​

0504, error 0504, XR200 0504, XR220 0504, Optima XR200amx error 0504, Optima XR220amx error 0504, Input stage: DC Bus rising time too slow during DC Bus charge [DC Bus
Question

What does error 0504 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0504: Input stage: DC Bus rising time too slow during DC Bus charge [DC Bus < 200V].

Possible Causes

Subsystem/category: Input Stage / DC Bus. Manual recovery/error class: ABORT.

Troubleshooting Steps

Check Power Control Limits and generator input/DC-bus diagnostics. Evaluate system battery state, input cabling and generator input-stage hardware.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 0505

Error Code System Message

​

0505, error 0505, XR200 0505, XR220 0505, Optima XR200amx error 0505, Optima XR220amx error 0505, Input stage: No input current detected during DC Bus charge
Question

What does error 0505 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0505: Input stage: No input current detected during DC Bus charge.

Possible Causes

Subsystem/category: Input Stage / DC Bus. Manual recovery/error class: ABORT.

Troubleshooting Steps

Check Power Control Limits and generator input/DC-bus diagnostics. Evaluate system battery state, input cabling and generator input-stage hardware.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 0510

Error Code System Message

​

0510, error 0510, XR200 0510, XR220 0510, Optima XR200amx error 0510, Optima XR220amx error 0510, Input stage: Over voltage on DC output detected [DC Bus > 460V]
Question

What does error 0510 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0510: Input stage: Over voltage on DC output detected [DC Bus > 460V].

Possible Causes

Subsystem/category: Input Stage / DC Bus. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Check Power Control Limits and generator input/DC-bus diagnostics. Evaluate system battery state, input cabling and generator input-stage hardware.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 0511

Error Code System Message

​

0511, error 0511, XR200 0511, XR220 0511, Optima XR200amx error 0511, Optima XR220amx error 0511, Input stage: Unexpected generator firmware error
Question

What does error 0511 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0511: Input stage: Unexpected generator firmware error.

Possible Causes

Subsystem/category: Input Stage / DC Bus. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Check Power Control Limits and generator input/DC-bus diagnostics. Evaluate system battery state, input cabling and generator input-stage hardware.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 0520

Error Code System Message

​

0520, error 0520, XR200 0520, XR220 0520, Optima XR200amx error 0520, Optima XR220amx error 0520, Input stage: Software over input current detected [Iin > 140A]
Question

What does error 0520 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0520: Input stage: Software over input current detected [Iin > 140A].

Possible Causes

Subsystem/category: Input Stage / DC Bus. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Check Power Control Limits and generator input/DC-bus diagnostics. Evaluate system battery state, input cabling and generator input-stage hardware.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

bottom of page