AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 102
Error Code System Message
102
Question
What should I check for errors 102 on the AMX-4 Series?
Short Answer
1) Perform recalibration of the “CAL BATT CHARGER” portion. 2) Replace the Battery Charger Board.
Possible Causes
Perform recalibration of the “CAL BATT CHARGER” portion. 2) Replace the Battery Charger Board.
Troubleshooting Steps
1) Perform recalibration of the “CAL BATT CHARGER” portion. 2) Replace the Battery Charger Board.
Parts Areas to Check
Battery Charger Board; Battery
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
Error 103
Error Code System Message
103
Question
What should I check for error 103 on the AMX-4 Series?
Short Answer
Safety contactor contacts may be stuck closed.
Possible Causes
Safety contactor contacts may be stuck closed.
Troubleshooting Steps
Safety contactor contacts may be stuck closed.
Parts Areas to Check
Safety Contactor
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
Error 103
Error Code System Message
ERROR 103 - Charge-current feedback high during trickle charge
Question
What does ERROR 103 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 103: Charge-current feedback high during trickle charge.
Possible Causes
Feedback exceeded expected trickle range.
Troubleshooting Steps
Check charger scale-select/feedback and actual trickle-current behavior.
Parts Areas to Check
Battery set; charger board; CPU board; voltage/current feedback wiring
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
Error 104
Error Code System Message
ERROR 104 - Charge-current feedback low during trickle charge
Question
What does ERROR 104 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 104: Charge-current feedback low during trickle charge.
Possible Causes
Feedback was below expected trickle range.
Troubleshooting Steps
Check charger scale-select/feedback and actual trickle-current behavior.
Parts Areas to Check
Battery set; charger board; CPU board; voltage/current feedback wiring
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
Error 104
Error Code System Message
104
Question
What should I check for error 104 on the AMX-4 Series?
Short Answer
Refer to Service Note 3684.
Possible Causes
Refer to Service Note 3684.
Troubleshooting Steps
Refer to Service Note 3684.
Parts Areas to Check