top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 0707

Error Code System Message

​

0707, error 0707, XR200 0707, XR220 0707, Optima XR200amx error 0707, Optima XR220amx error 0707, Thermal: Power board temperature is rising
Question

What does error 0707 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0707: Thermal: Power board temperature is rising.

Possible Causes

Subsystem/category: Thermal. Manual recovery/error class: DEBUG.

Troubleshooting Steps

Allow cooling when appropriate, verify airflow, and run generator thermal/70C diagnostics. Check the board or sensor identified by the code.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 0708

Error Code System Message

​

0708, error 0708, XR200 0708, XR220 0708, Optima XR200amx error 0708, Optima XR220amx error 0708, Thermal: Filament board temperature is rising
Question

What does error 0708 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0708: Thermal: Filament board temperature is rising.

Possible Causes

Subsystem/category: Thermal. Manual recovery/error class: DEBUG.

Troubleshooting Steps

Allow cooling when appropriate, verify airflow, and run generator thermal/70C diagnostics. Check the board or sensor identified by the code.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 0709

Error Code System Message

​

0709, error 0709, XR200 0709, XR220 0709, Optima XR200amx error 0709, Optima XR220amx error 0709, Thermal: Rotor board temperature is rising
Question

What does error 0709 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0709: Thermal: Rotor board temperature is rising.

Possible Causes

Subsystem/category: Thermal. Manual recovery/error class: DEBUG.

Troubleshooting Steps

Allow cooling when appropriate, verify airflow, and run generator thermal/70C diagnostics. Check the board or sensor identified by the code.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 0782

Error Code System Message

​

0782, error 0782, XR200 0782, XR220 0782, Optima XR200amx error 0782, Optima XR220amx error 0782, Thermal: Tank temperature sensor failure is detected
Question

What does error 0782 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0782: Thermal: Tank temperature sensor failure is detected.

Possible Causes

Subsystem/category: Thermal. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Allow cooling when appropriate, verify airflow, and run generator thermal/70C diagnostics. Check the board or sensor identified by the code.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 0783

Error Code System Message

​

0783, error 0783, XR200 0783, XR220 0783, Optima XR200amx error 0783, Optima XR220amx error 0783, Thermal: Power board temperature sensor failure
Question

What does error 0783 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0783: Thermal: Power board temperature sensor failure.

Possible Causes

Subsystem/category: Thermal. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Allow cooling when appropriate, verify airflow, and run generator thermal/70C diagnostics. Check the board or sensor identified by the code.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

bottom of page