AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 105
Error Code System Message
ERROR 105 - Charger open too long / full-current condition persists
Question
What does ERROR 105 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 105: Charger open too long / full-current condition persists.
Possible Causes
Charger remained open/full output longer than expected.
Troubleshooting Steps
Verify charger calibration, battery voltage/SOC and current feedback; inspect charger-board operation.
Parts Areas to Check
Battery set; charger board; CPU board; voltage/current feedback wiring
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
Error 105
Error Code System Message
105
Question
What should I check for errors 105 on the AMX-4 Series?
Short Answer
1) Perform recalibration of the “CAL BATT CHARGER” portion. 2) Replace the Battery Charger Board.
Possible Causes
Perform recalibration of the “CAL BATT CHARGER” portion. Replace the Battery Charger Board.
Troubleshooting Steps
1) Perform recalibration of the “CAL BATT CHARGER” portion. 2) Replace the Battery Charger Board.
Parts Areas to Check
Battery Charger Board; Battery
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
Error 106
Error Code System Message
106
Question
What should I check for errors 106 on the AMX-4 Series?
Short Answer
1) Perform recalibration of the “CAL BATT CHARGER” portion. 2) Replace the Battery Charger Board.
Possible Causes
Perform recalibration of the “CAL BATT CHARGER” portion. Replace the Battery Charger Board.
Troubleshooting Steps
1) Perform recalibration of the “CAL BATT CHARGER” portion. 2) Replace the Battery Charger Board.
Parts Areas to Check
Battery Charger Board; Battery
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
Error 106
Error Code System Message
ERROR 106 - Charge-current feedback zero
Question
What does ERROR 106 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 106: Charge-current feedback zero.
Possible Causes
No charging-current feedback detected.
Troubleshooting Steps
Check charger current feedback, charger board, CPU connection and charging path.
Parts Areas to Check
Battery set; charger board; CPU board; voltage/current feedback wiring
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
Error 107
Error Code System Message
ERROR 107 - Charge-clamp voltage exceeded at zero current
Question
What does ERROR 107 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 107: Charge-clamp voltage exceeded at zero current.
Possible Causes
Charge-clamp voltage exceeded the allowed condition.
Troubleshooting Steps
Check charger scale-select/feedback and charger-board output.
Parts Areas to Check
Battery set; charger board; CPU board; voltage/current feedback wiring