AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 105

Error Code System Message

ERROR 105 - Charger open too long / full-current condition persists

ERROR 105, Charger open too long / full-current condition persists, AMX-4 ERROR 105, AMX-4+ ERROR 105, Definium AMX 700 ERROR 105, GE AMX error ERROR 105
Question

What does ERROR 105 mean on a GE AMX-Series mobile X-ray?

Short Answer

ERROR 105: Charger open too long / full-current condition persists.

Possible Causes

Charger remained open/full output longer than expected.

Troubleshooting Steps

Verify charger calibration, battery voltage/SOC and current feedback; inspect charger-board operation.

Parts Areas to Check

Battery set; charger board; CPU board; voltage/current feedback wiring

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 105

Error Code System Message

105

AMX-4, AMX-4+, Definium AMX 700, ERROR 105
Question

What should I check for errors 105 on the AMX-4 Series?

Short Answer

1) Perform recalibration of the “CAL BATT CHARGER” portion. 2) Replace the Battery Charger Board.

Possible Causes

Perform recalibration of the “CAL BATT CHARGER” portion. Replace the Battery Charger Board.

Troubleshooting Steps

1) Perform recalibration of the “CAL BATT CHARGER” portion. 2) Replace the Battery Charger Board.

Parts Areas to Check

Battery Charger Board; Battery

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 106

Error Code System Message

106

AMX-4, AMX-4+, Definium AMX 700, ERROR 106
Question

What should I check for errors 106 on the AMX-4 Series?

Short Answer

1) Perform recalibration of the “CAL BATT CHARGER” portion. 2) Replace the Battery Charger Board.

Possible Causes

Perform recalibration of the “CAL BATT CHARGER” portion. Replace the Battery Charger Board.

Troubleshooting Steps

1) Perform recalibration of the “CAL BATT CHARGER” portion. 2) Replace the Battery Charger Board.

Parts Areas to Check

Battery Charger Board; Battery

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 106

Error Code System Message

ERROR 106 - Charge-current feedback zero

ERROR 106, Charge-current feedback zero, AMX-4 ERROR 106, AMX-4+ ERROR 106, Definium AMX 700 ERROR 106, GE AMX error ERROR 106
Question

What does ERROR 106 mean on a GE AMX-Series mobile X-ray?

Short Answer

ERROR 106: Charge-current feedback zero.

Possible Causes

No charging-current feedback detected.

Troubleshooting Steps

Check charger current feedback, charger board, CPU connection and charging path.

Parts Areas to Check

Battery set; charger board; CPU board; voltage/current feedback wiring

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 107

Error Code System Message

ERROR 107 - Charge-clamp voltage exceeded at zero current

ERROR 107, Charge-clamp voltage exceeded at zero current, AMX-4 ERROR 107, AMX-4+ ERROR 107, Definium AMX 700 ERROR 107, GE AMX error ERROR 107
Question

What does ERROR 107 mean on a GE AMX-Series mobile X-ray?

Short Answer

ERROR 107: Charge-clamp voltage exceeded at zero current.

Possible Causes

Charge-clamp voltage exceeded the allowed condition.

Troubleshooting Steps

Check charger scale-select/feedback and charger-board output.

Parts Areas to Check

Battery set; charger board; CPU board; voltage/current feedback wiring

For Related Models

AMX-4, AMX-4+, Definium AMX 700