AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 0B84
Error Code System Message
Question
What does error 0B84 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0B84: mA: mA measurement at start of exposure lower than expected.
Possible Causes
Subsystem/category: mA Generation. Manual recovery/error class: DEBUG.
Troubleshooting Steps
Review related generator errors. Run mA Measurement and the applicable filament-driver tests; inspect tube/filament connections and use the specific error workflow before replacing parts.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 0B85
Error Code System Message
Question
What does error 0B85 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0B85: mA: mA measurement at start of exposure higher than expected.
Possible Causes
Subsystem/category: mA Generation. Manual recovery/error class: DEBUG.
Troubleshooting Steps
Review related generator errors. Run mA Measurement and the applicable filament-driver tests; inspect tube/filament connections and use the specific error workflow before replacing parts.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 0B86
Error Code System Message
Question
What does error 0B86 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0B86: mA: Filament ageing saturation detected when computed target reach command limit.
Possible Causes
Subsystem/category: mA Generation. Manual recovery/error class: DEBUG.
Troubleshooting Steps
Review related generator errors. Run mA Measurement and the applicable filament-driver tests; inspect tube/filament connections and use the specific error workflow before replacing parts.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 0B87
Error Code System Message
Question
What does error 0B87 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0B87: mA: Measured mA is out of range during exposure [mA < 10% of target].
Possible Causes
Subsystem/category: mA Generation. Manual recovery/error class: ABORT.
Troubleshooting Steps
Review related generator errors. Run mA Measurement and the applicable filament-driver tests; inspect tube/filament connections and use the specific error workflow before replacing parts.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 0B8x, mA errors
Error Code System Message
Question
How should I approach XR200amx/XR220amx 0B8x mA errors?
Short Answer
0B8x codes report mA-measurement and filament-aging conditions during generator operation and exposure.
Possible Causes
Filament calibration/aging; mA measurement path; filament driver; tube problem.
Troubleshooting Steps
Review the exact 0B8x code, run mA Measurement and the filament-driver tests, and perform the approved filament calibration/error-specific procedure where directed.
Parts Areas to Check