top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 0B84

Error Code System Message

​

0B84, error 0B84, XR200 0B84, XR220 0B84, Optima XR200amx error 0B84, Optima XR220amx error 0B84, mA: mA measurement at start of exposure lower than expected
Question

What does error 0B84 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0B84: mA: mA measurement at start of exposure lower than expected.

Possible Causes

Subsystem/category: mA Generation. Manual recovery/error class: DEBUG.

Troubleshooting Steps

Review related generator errors. Run mA Measurement and the applicable filament-driver tests; inspect tube/filament connections and use the specific error workflow before replacing parts.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 0B85

Error Code System Message

​

0B85, error 0B85, XR200 0B85, XR220 0B85, Optima XR200amx error 0B85, Optima XR220amx error 0B85, mA: mA measurement at start of exposure higher than expected
Question

What does error 0B85 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0B85: mA: mA measurement at start of exposure higher than expected.

Possible Causes

Subsystem/category: mA Generation. Manual recovery/error class: DEBUG.

Troubleshooting Steps

Review related generator errors. Run mA Measurement and the applicable filament-driver tests; inspect tube/filament connections and use the specific error workflow before replacing parts.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 0B86

Error Code System Message

​

0B86, error 0B86, XR200 0B86, XR220 0B86, Optima XR200amx error 0B86, Optima XR220amx error 0B86, mA: Filament ageing saturation detected when computed target reach command limit
Question

What does error 0B86 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0B86: mA: Filament ageing saturation detected when computed target reach command limit.

Possible Causes

Subsystem/category: mA Generation. Manual recovery/error class: DEBUG.

Troubleshooting Steps

Review related generator errors. Run mA Measurement and the applicable filament-driver tests; inspect tube/filament connections and use the specific error workflow before replacing parts.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 0B87

Error Code System Message

​

0B87, error 0B87, XR200 0B87, XR220 0B87, Optima XR200amx error 0B87, Optima XR220amx error 0B87, mA: Measured mA is out of range during exposure [mA
Question

What does error 0B87 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0B87: mA: Measured mA is out of range during exposure [mA < 10% of target].

Possible Causes

Subsystem/category: mA Generation. Manual recovery/error class: ABORT.

Troubleshooting Steps

Review related generator errors. Run mA Measurement and the applicable filament-driver tests; inspect tube/filament connections and use the specific error workflow before replacing parts.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 0B8x, mA errors

Error Code System Message

​

XR220 0B84, XR200 mA error, filament aging 0B86
Question

How should I approach XR200amx/XR220amx 0B8x mA errors?

Short Answer

0B8x codes report mA-measurement and filament-aging conditions during generator operation and exposure.

Possible Causes

Filament calibration/aging; mA measurement path; filament driver; tube problem.

Troubleshooting Steps

Review the exact 0B8x code, run mA Measurement and the filament-driver tests, and perform the approved filament calibration/error-specific procedure where directed.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

bottom of page