AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 3600132
Error Code System Message
Question
What does error 3600132 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 3600132: Generator Debug Error.
Possible Causes
Subsystem/category: Generator. Manual recovery/error class: DEBUG.
Troubleshooting Steps
Underlying Djinn DEBUG-class condition is present.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 3600133
Error Code System Message
Question
What does error 3600133 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 3600133: Generator Inhibit thermal Error.
Possible Causes
Subsystem/category: Generator. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
A detailed thermal Djinn error is inhibiting generator operation.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 3600133
Error Code System Message
Question
What does system error 3600133 mean on XR200amx or XR220amx?
Short Answer
3600133 is the generic Generator Inhibit Thermal Error.
Possible Causes
A thermal generator condition is inhibiting operation.
Troubleshooting Steps
Check the detailed Djinn thermal error code, allow cooling where appropriate, verify airflow and run the applicable thermal diagnostic before replacing parts.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 3600134
Error Code System Message
Question
What does error 3600134 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 3600134: Generator Inhibit non-thermal Error.
Possible Causes
Subsystem/category: Generator. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
A detailed non-thermal Djinn error is inhibiting generator operation.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 3600134
Error Code System Message
Question
What does system error 3600134 mean on XR200amx or XR220amx?
Short Answer
3600134 is the generic Generator Inhibit Non-Thermal Error.
Possible Causes
A non-thermal Djinn condition is inhibiting operation.
Troubleshooting Steps
Identify the detailed Djinn code in the Class M log and troubleshoot the indicated filament, rotor, DC-bus, kV, system, boot or communication subsystem.
Parts Areas to Check