top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 3600132

Error Code System Message

​

3600132, error 3600132, XR200 3600132, XR220 3600132, Optima XR200amx error 3600132, Optima XR220amx error 3600132, Generator Debug Error
Question

What does error 3600132 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 3600132: Generator Debug Error.

Possible Causes

Subsystem/category: Generator. Manual recovery/error class: DEBUG.

Troubleshooting Steps

Underlying Djinn DEBUG-class condition is present.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 3600133

Error Code System Message

​

3600133, error 3600133, XR200 3600133, XR220 3600133, Optima XR200amx error 3600133, Optima XR220amx error 3600133, Generator Inhibit thermal Error
Question

What does error 3600133 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 3600133: Generator Inhibit thermal Error.

Possible Causes

Subsystem/category: Generator. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

A detailed thermal Djinn error is inhibiting generator operation.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 3600133

Error Code System Message

​

3600133 thermal inhibit, XR220 generator overheat
Question

What does system error 3600133 mean on XR200amx or XR220amx?

Short Answer

3600133 is the generic Generator Inhibit Thermal Error.

Possible Causes

A thermal generator condition is inhibiting operation.

Troubleshooting Steps

Check the detailed Djinn thermal error code, allow cooling where appropriate, verify airflow and run the applicable thermal diagnostic before replacing parts.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 3600134

Error Code System Message

​

3600134, error 3600134, XR200 3600134, XR220 3600134, Optima XR200amx error 3600134, Optima XR220amx error 3600134, Generator Inhibit non-thermal Error
Question

What does error 3600134 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 3600134: Generator Inhibit non-thermal Error.

Possible Causes

Subsystem/category: Generator. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

A detailed non-thermal Djinn error is inhibiting generator operation.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 3600134

Error Code System Message

​

3600134 generator inhibit, XR200 Djinn error
Question

What does system error 3600134 mean on XR200amx or XR220amx?

Short Answer

3600134 is the generic Generator Inhibit Non-Thermal Error.

Possible Causes

A non-thermal Djinn condition is inhibiting operation.

Troubleshooting Steps

Identify the detailed Djinn code in the Class M log and troubleshoot the indicated filament, rotor, DC-bus, kV, system, boot or communication subsystem.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

bottom of page