AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 8301070
Error Code System Message
Question
What does error 8301070 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 8301070: Battery stack voltage over 180V – charge inhibited.
Possible Causes
Subsystem/category: System Battery. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Evaluate pack SOC mismatch, charging control and connections.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 8301071
Error Code System Message
Question
What does error 8301071 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 8301071: Battery pack 1 voltage over 50V – breaker to trip.
Possible Causes
Subsystem/category: System Battery. Manual recovery/error class: TRIP.
Troubleshooting Steps
Confirm whether the overvoltage follows pack 1 when pack positions are swapped; evaluate charging controls.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 8301072
Error Code System Message
Question
What does error 8301072 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 8301072: Battery pack 2 voltage over 50V – breaker to trip.
Possible Causes
Subsystem/category: System Battery. Manual recovery/error class: TRIP.
Troubleshooting Steps
Confirm whether the overvoltage follows pack 2 when pack positions are swapped; evaluate charging controls.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 8301073
Error Code System Message
Question
What does error 8301073 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 8301073: Battery pack 3 voltage over 50V – breaker to trip.
Possible Causes
Subsystem/category: System Battery. Manual recovery/error class: TRIP.
Troubleshooting Steps
Confirm whether the overvoltage follows pack 3 when pack positions are swapped; evaluate charging controls.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 8301074
Error Code System Message
Question
What does error 8301074 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 8301074: Battery pack 4 voltage over 50V – breaker to trip.
Possible Causes
Subsystem/category: System Battery. Manual recovery/error class: TRIP.
Troubleshooting Steps
Confirm whether the overvoltage follows pack 4 when pack positions are swapped; evaluate charging controls.
Parts Areas to Check