Error 8301070

Error Code System Message

​

8301070, error 8301070, XR200 8301070, XR220 8301070, Optima XR200amx error 8301070, Optima XR220amx error 8301070, Battery stack voltage over 180V – charge inhibited
Question

What does error 8301070 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 8301070: Battery stack voltage over 180V – charge inhibited.

Possible Causes

Subsystem/category: System Battery. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Evaluate pack SOC mismatch, charging control and connections.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 8301071

Error Code System Message

​

8301071, error 8301071, XR200 8301071, XR220 8301071, Optima XR200amx error 8301071, Optima XR220amx error 8301071, Battery pack 1 voltage over 50V – breaker to trip
Question

What does error 8301071 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 8301071: Battery pack 1 voltage over 50V – breaker to trip.

Possible Causes

Subsystem/category: System Battery. Manual recovery/error class: TRIP.

Troubleshooting Steps

Confirm whether the overvoltage follows pack 1 when pack positions are swapped; evaluate charging controls.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 8301072

Error Code System Message

​

8301072, error 8301072, XR200 8301072, XR220 8301072, Optima XR200amx error 8301072, Optima XR220amx error 8301072, Battery pack 2 voltage over 50V – breaker to trip
Question

What does error 8301072 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 8301072: Battery pack 2 voltage over 50V – breaker to trip.

Possible Causes

Subsystem/category: System Battery. Manual recovery/error class: TRIP.

Troubleshooting Steps

Confirm whether the overvoltage follows pack 2 when pack positions are swapped; evaluate charging controls.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 8301073

Error Code System Message

​

8301073, error 8301073, XR200 8301073, XR220 8301073, Optima XR200amx error 8301073, Optima XR220amx error 8301073, Battery pack 3 voltage over 50V – breaker to trip
Question

What does error 8301073 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 8301073: Battery pack 3 voltage over 50V – breaker to trip.

Possible Causes

Subsystem/category: System Battery. Manual recovery/error class: TRIP.

Troubleshooting Steps

Confirm whether the overvoltage follows pack 3 when pack positions are swapped; evaluate charging controls.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 8301074

Error Code System Message

​

8301074, error 8301074, XR200 8301074, XR220 8301074, Optima XR200amx error 8301074, Optima XR220amx error 8301074, Battery pack 4 voltage over 50V – breaker to trip
Question

What does error 8301074 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 8301074: Battery pack 4 voltage over 50V – breaker to trip.

Possible Causes

Subsystem/category: System Battery. Manual recovery/error class: TRIP.

Troubleshooting Steps

Confirm whether the overvoltage follows pack 4 when pack positions are swapped; evaluate charging controls.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx