top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Wireless detector issues

Error Code System Message

​

XR220 wireless detector, UWB problem, FlashPad signal strength
Question

Which wireless detector diagnostics are available on XR220amx and digitally upgraded XR200amx?

Short Answer

Wireless diagnostics include Wireless Comm, Signal Strength, Battery Health, BIST, Ping, Ethernet Stats, temperature and detector power-supply tests.

Possible Causes

UWB authentication/signal problem; detector battery condition; detector internal fault; data-transfer issue.

Troubleshooting Steps

Use Wireless Comm to identify the connection stage and signal condition. Check Battery Health, then BIST and Ping. Run Ethernet Stats for transfer problems.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Detector keep losing its wireless AP connection

Error Code System Message

​

XR240 detector disconnects, FlashPad loses wireless, AP connection drops
Question

Why does my XR240amx detector keep losing its wireless AP connection?

Short Answer

Check country/channel compatibility, AP power, detector distance, signal quality, and whether the detector received a valid IP address.

Possible Causes

Country/channel mismatch; loss of AP 12 VDC; detector out of range; low signal; heavy interference; invalid detector IP.

Troubleshooting Steps

Verify country configuration and AP power. Move the detector closer to the system/AP. Review connected clients and detector IP assignment. Reboot detector and system after correcting configuration or power issues.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Errors 0403 or 0413

Error Code System Message

​

XR240 0403, 0413, rotor overcurrent, tube rotor impedance
Question

What does XR240amx generator error 0403 or 0413 mean?

Short Answer

These generator errors indicate overcurrent detected on the rotation phases.

Possible Causes

Rotor circuit fault; tube stator/wiring issue; Main Auxiliary module issue.

Troubleshooting Steps

Run the Tube Rotor Impedance diagnostic and applicable stator resistance checks. If tube stator resistance is incorrect, the manual directs replacement of the tube; if resistance is correct, inspect rotor harness wiring and Main Auxiliary circuitry.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

mA measurement lower than expected

Error Code System Message

​

XR240 mA low, error 0B84, error 0B87, filament calibration
Question

How do I troubleshoot XR240amx mA measurement lower than expected?

Short Answer

The manual recommends performing a new filament calibration first and then continuing with filament-driver diagnostics if the error persists.

Possible Causes

Filament calibration/aging; filament driver issue; measurement path issue; tube issue.

Troubleshooting Steps

Run Filament Drive Calibration. During calibration, some low/high mA messages can occur and may be acceptable if calibration completes successfully. If the problem persists afterward, follow the filament driver and mA measurement diagnostic sequence.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Access point boot failed

Error Code System Message

​

XR240 access point boot failed, AP not booting, wireless access point error
Question

What should I do when the XR240amx says “Access point boot failed”?

Short Answer

Allow the AP boot attempt to finish; if it fails, the manual says to reboot and then verify country configuration, AP connections, power, and firmware.

Possible Causes

AP not finished booting; country configuration issue; Ethernet/power connection problem; AP hardware fault.

Troubleshooting Steps

Wait up to about five minutes for AP boot. If the boot failure message appears, reboot the system. Recommit country configuration, inspect AP Ethernet and power connections, and continue AP diagnostics if failure returns.

Parts Areas to Check

​

For Related Models

Optima XR240amx

bottom of page