AMX Troubleshooting & Technical Support

Results Count

Error Number

Detector connect to the AP but not get a proper IP address

Error Code System Message

​

XR240 detector no IP, FlashPad IP address, AP clients
Question

Why does an XR240amx detector connect to the AP but not get a proper IP address?

Short Answer

Use the AP client list to verify the detector is recognized and has an assigned IP address, then reboot the detector and system if needed.

Possible Causes

Unrecognized detector connection; failed IP assignment; AP pairing state issue.

Troubleshooting Steps

Run the AP clients diagnostic, confirm the detector is listed and has a proper IP, then reboot the detector and system. If AP power was lost, detectors may need to be paired again.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Detector fails to boot wirelessly

Error Code System Message

​

XR240 detector wireless boot failure, FlashPad not booting, detector not primary
Question

Why does my XR240amx detector fail to boot wirelessly?

Short Answer

Confirm the detector is registered, paired, selected as primary, and can boot successfully on the tether connection.

Possible Causes

Detector not registered; too many detectors registered; detector not primary; detector not paired; wireless communication failure.

Troubleshooting Steps

Verify detector registration and primary status. Pair the detector in the charging bin. Attempt a tethered boot to isolate the fault to wireless communication. If tether boot succeeds but wireless fails, continue AP/wireless diagnostics and collect a system snapshot if needed.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Detector has dropped

Error Code System Message

​

dropped FlashPad, XR240 detector impact, detector QAP after drop
Question

What should I do if an XR240amx detector is dropped?

Short Answer

The manual states that a QAP should be performed and passed before imaging is attempted after a drop or impact.

Possible Causes

Mechanical impact or internal detector damage.

Troubleshooting Steps

Remove the detector from clinical use until the required quality-assurance procedure (QAP) is completed and passes. Inspect the detector for physical damage and escalate if the QAP fails.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Detector is not primary

Error Code System Message

​

XR240 primary detector, set FlashPad primary, detector connected not active
Question

How do I make a detector primary on the GE Optima XR240amx?

Short Answer

Select the desired detector as the primary detector in Detector Management; a reboot may be required for boot to complete.

Possible Causes

Wrong detector selected as primary.

Troubleshooting Steps

Open Detector Management, select the desired detector as primary, and allow the detector to complete boot. If necessary, reboot the detector after changing primary status.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Detector is still paired to another system

Error Code System Message

​

XR240 detector paired to different system, re-pair FlashPad, detector swap
Question

How do I fix an XR240amx detector that is paired to another system?

Short Answer

Review registered/paired detectors and insert the detector into the intended system’s bin to auto-pair it.

Possible Causes

Detector is still paired with another XR240 system.

Troubleshooting Steps

Check the registered-detector list and current primary/secondary assignments. Insert the detector into the intended system’s bin so the pairing process can run.

Parts Areas to Check

​

For Related Models

Optima XR240amx