Error 0204
Error Code System Message
Question
What does error 0204 mean on a GE Optima XR240amx?
Short Answer
Error 0204: Filament drive current below minimum limit.
Possible Causes
Subsystem/category: Fil Large. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0205
Error Code System Message
Question
What does error 0205 mean on a GE Optima XR240amx?
Short Answer
Error 0205: Filament inverter is off and current is not zero.
Possible Causes
Subsystem/category: Fil Large. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0207
Error Code System Message
Question
What does error 0207 mean on a GE Optima XR240amx?
Short Answer
Error 0207: Overcurrent detected on filament.
Possible Causes
Subsystem/category: Fil Large. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0208
Error Code System Message
Question
What does error 0208 mean on a GE Optima XR240amx?
Short Answer
Error 0208: 48 V filament supply circuit out of range (44–52 V).
Possible Causes
Subsystem/category: Fil Large. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0281
Error Code System Message
Question
What does error 0281 mean on a GE Optima XR240amx?
Short Answer
Error 0281: Unexpected generator firmware error.
Possible Causes
Subsystem/category: Fil Small. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts.
Parts Areas to Check