AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 0523
Error Code System Message
Question
What does error 0523 mean on a GE Optima XR240amx?
Short Answer
Error 0523: Input stage damaged.
Possible Causes
Subsystem/category: Input Stage. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review related generator and system-battery errors. Run Power Supplies / Power Control Limits and the applicable input-voltage or DC-bus diagnostics. Inspect the generator input-power path.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0524
Error Code System Message
Question
What does error 0524 mean on a GE Optima XR240amx?
Short Answer
Error 0524: Hardware over-input current detected (Iin >140 A).
Possible Causes
Subsystem/category: Input Stage. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review related generator and system-battery errors. Run Power Supplies / Power Control Limits and the applicable input-voltage or DC-bus diagnostics. Inspect the generator input-power path.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0525
Error Code System Message
Question
What does error 0525 mean on a GE Optima XR240amx?
Short Answer
Error 0525: Boot timeout when boot is started.
Possible Causes
Subsystem/category: Input Stage. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review related generator and system-battery errors. Run Power Supplies / Power Control Limits and the applicable input-voltage or DC-bus diagnostics. Inspect the generator input-power path.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0580
Error Code System Message
Question
What does error 0580 mean on a GE Optima XR240amx?
Short Answer
Error 0580: DC-bus timeout when boot is started.
Possible Causes
Subsystem/category: Input Stage. Manual recovery/error class: ABORT.
Troubleshooting Steps
Review related generator and system-battery errors. Run Power Supplies / Power Control Limits and the applicable input-voltage or DC-bus diagnostics. Inspect the generator input-power path.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0581
Error Code System Message
Question
What does error 0581 mean on a GE Optima XR240amx?
Short Answer
Error 0581: Low input voltage during boost.
Possible Causes
Subsystem/category: Input Stage. Manual recovery/error class: DEBUG.
Troubleshooting Steps
Review related generator and system-battery errors. Run Power Supplies / Power Control Limits and the applicable input-voltage or DC-bus diagnostics. Inspect the generator input-power path.