Error Number

Error 0604

Error Code System Message

​

0604, error 0604, XR240 0604, Optima XR240amx error 0604, Maximum kV reached on anode (>125% of anode kV target)
Question

What does error 0604 mean on a GE Optima XR240amx?

Short Answer

Error 0604: Maximum kV reached on anode (>125% of anode kV target).

Possible Causes

Subsystem/category: kV. Manual recovery/error class: ABORT.

Troubleshooting Steps

Review related generator errors. Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics. Inspect the HV/tube cable path only under qualified service procedures.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Error 0605

Error Code System Message

​

0605, error 0605, XR240 0605, Optima XR240amx error 0605, Maximum kV reached on cathode (>125% of cathode kV target)
Question

What does error 0605 mean on a GE Optima XR240amx?

Short Answer

Error 0605: Maximum kV reached on cathode (>125% of cathode kV target).

Possible Causes

Subsystem/category: kV. Manual recovery/error class: ABORT.

Troubleshooting Steps

Review related generator errors. Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics. Inspect the HV/tube cable path only under qualified service procedures.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Error 0606

Error Code System Message

​

0606, error 0606, XR240 0606, Optima XR240amx error 0606, Too many spits detected (>8 per second)
Question

What does error 0606 mean on a GE Optima XR240amx?

Short Answer

Error 0606: Too many spits detected (>8 per second).

Possible Causes

Subsystem/category: kV. Manual recovery/error class: ABORT.

Troubleshooting Steps

Review related generator errors. Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics. Inspect the HV/tube cable path only under qualified service procedures.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Error 0607

Error Code System Message

​

0607, error 0607, XR240 0607, Optima XR240amx error 0607, Inverter current mismatch between measured and expected values
Question

What does error 0607 mean on a GE Optima XR240amx?

Short Answer

Error 0607: Inverter current mismatch between measured and expected values.

Possible Causes

Subsystem/category: kV. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Review related generator errors. Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics. Inspect the HV/tube cable path only under qualified service procedures.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Error 0608

Error Code System Message

​

0608, error 0608, XR240 0608, Optima XR240amx error 0608, Inverter overcurrent detected (ILR >250 A)
Question

What does error 0608 mean on a GE Optima XR240amx?

Short Answer

Error 0608: Inverter overcurrent detected (ILR >250 A).

Possible Causes

Subsystem/category: kV. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Review related generator errors. Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics. Inspect the HV/tube cable path only under qualified service procedures.

Parts Areas to Check

​

For Related Models

Optima XR240amx

bottom of page