AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 0604
Error Code System Message
Question
What does error 0604 mean on a GE Optima XR240amx?
Short Answer
Error 0604: Maximum kV reached on anode (>125% of anode kV target).
Possible Causes
Subsystem/category: kV. Manual recovery/error class: ABORT.
Troubleshooting Steps
Review related generator errors. Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics. Inspect the HV/tube cable path only under qualified service procedures.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0605
Error Code System Message
Question
What does error 0605 mean on a GE Optima XR240amx?
Short Answer
Error 0605: Maximum kV reached on cathode (>125% of cathode kV target).
Possible Causes
Subsystem/category: kV. Manual recovery/error class: ABORT.
Troubleshooting Steps
Review related generator errors. Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics. Inspect the HV/tube cable path only under qualified service procedures.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0606
Error Code System Message
Question
What does error 0606 mean on a GE Optima XR240amx?
Short Answer
Error 0606: Too many spits detected (>8 per second).
Possible Causes
Subsystem/category: kV. Manual recovery/error class: ABORT.
Troubleshooting Steps
Review related generator errors. Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics. Inspect the HV/tube cable path only under qualified service procedures.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0607
Error Code System Message
Question
What does error 0607 mean on a GE Optima XR240amx?
Short Answer
Error 0607: Inverter current mismatch between measured and expected values.
Possible Causes
Subsystem/category: kV. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review related generator errors. Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics. Inspect the HV/tube cable path only under qualified service procedures.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0608
Error Code System Message
Question
What does error 0608 mean on a GE Optima XR240amx?
Short Answer
Error 0608: Inverter overcurrent detected (ILR >250 A).
Possible Causes
Subsystem/category: kV. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review related generator errors. Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics. Inspect the HV/tube cable path only under qualified service procedures.
Parts Areas to Check