AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 218

Error Code System Message

ERROR 218 - Handle signals out of range

ERROR 218, Handle signals out of range, AMX-4 ERROR 218, AMX-4+ ERROR 218, Definium AMX 700 ERROR 218, GE AMX error ERROR 218
Question

What does ERROR 218 mean on a GE AMX-Series mobile X-ray?

Short Answer

ERROR 218: Handle signals out of range.

Possible Causes

One or both drive-handle signals were outside calibration range.

Troubleshooting Steps

Calibrate handle; inspect Hall-effect sensors, handle mechanics and CPU inputs.

Parts Areas to Check

Drive control board; drive handle/Hall sensors; motor relays; drive motors; feedback wiring

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 23

Error Code System Message

ERROR 23 - Display status signal low or invalid

ERROR 23, Display status signal low or invalid, AMX-4 ERROR 23, AMX-4+ ERROR 23, Definium AMX 700 ERROR 23, GE AMX error ERROR 23
Question

What does ERROR 23 mean on a GE AMX-Series mobile X-ray?

Short Answer

ERROR 23: Display status signal low or invalid.

Possible Causes

DISPLAY OK/status line indicated a display/controller problem.

Troubleshooting Steps

Check display-status signal, display controller and related CPU/display connections.

Parts Areas to Check

CPU board; calibration data; subsystem identified by the error

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 320

Error Code System Message

ERROR 320 - 60 Hz inverter feedback missing at field-light command

ERROR 320, 60 Hz inverter feedback missing at field-light command, AMX-4 ERROR 320, AMX-4+ ERROR 320, Definium AMX 700 ERROR 320, GE AMX error ERROR 320
Question

What does ERROR 320 mean on a GE AMX-Series mobile X-ray?

Short Answer

ERROR 320: 60 Hz inverter feedback missing at field-light command.

Possible Causes

60 Hz inverter feedback did not appear when field light was commanded.

Troubleshooting Steps

Check 60 Hz enable/relay/inverter signals and Generator Control 1 connections.

Parts Areas to Check

Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 320

Error Code System Message

320

AMX-4, AMX-4+, Definium AMX 700, ERROR 320, Error 320
Question

What should I check for error 320 on the AMX-4 Series?

Short Answer

Note: (And Possibly 458 Error) 1) Check fuse F93 on Rotor Control Board 2) Check fuse F7 on AMX1 A3 3) Check for 60 Hz phase 1 signal at Anode of CR53 and 60 Hz phase 2 signal at Anode of CR66 on Rotor Control Board 4) Defective Rotor Control Board 5) Defective CPU Board 6) Check Stator/Rotor cable 7) Bad CR179 on Filament kVp Control Board

Possible Causes

Note: (And Possibly 458 Error) 1) Check fuse F93 on Rotor Control Board 2) Check fuse F7 on AMX1 A3 3) Check for 60 Hz phase 1 signal at Anode of CR53 and 60 Hz phase 2 signal at Anode of CR66 on Rotor Control Board 4) Defective Rotor Control Board 5) Defective CPU Board 6) Check Stator/Rotor cable 7) Bad CR179 on Filament kVp Control Board. 8) Bad Rotor Controller Board if overheating, fuse blown, circuit breaker kept tripping, bad 1 kHz board

Troubleshooting Steps

Note: (And Possibly 458 Error) 1) Check fuse F93 on Rotor Control Board 2) Check fuse F7 on AMX1 A3 3) Check for 60 Hz phase 1 signal at Anode of CR53 and 60 Hz phase 2 signal at Anode of CR66 on Rotor Control Board 4) Defective Rotor Control Board 5) Defective CPU Board 6) Check Stator/Rotor cable 7) Bad CR179 on Filament kVp Control Board

Parts Areas to Check

CPU Board; Rotor Control Board; Stator; Rotor

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 321

Error Code System Message

ERROR 321 - 60 Hz inverter feedback missing after field light off

ERROR 321, 60 Hz inverter feedback missing after field light off, AMX-4 ERROR 321, AMX-4+ ERROR 321, Definium AMX 700 ERROR 321, GE AMX error ERROR 321
Question

What does ERROR 321 mean on a GE AMX-Series mobile X-ray?

Short Answer

ERROR 321: 60 Hz inverter feedback missing after field light off.

Possible Causes

60 Hz inverter feedback failed after the field lamp was turned off.

Troubleshooting Steps

Check 60 Hz enable/relay/inverter signal path and generator-control board.

Parts Areas to Check

Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault

For Related Models

AMX-4, AMX-4+, Definium AMX 700