top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 0B82

Error Code System Message

​

0B82, error 0B82, XR240 0B82, Optima XR240amx error 0B82, No mAs measured after exposure start
Question

What does error 0B82 mean on a GE Optima XR240amx?

Short Answer

Error 0B82: No mAs measured after exposure start.

Possible Causes

Subsystem/category: mA. Manual recovery/error class: ABORT.

Troubleshooting Steps

Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Error 0B83

Error Code System Message

​

0B83, error 0B83, XR240 0B83, Optima XR240amx error 0B83, Inconsistency between mA anode and mA cathode measurement
Question

What does error 0B83 mean on a GE Optima XR240amx?

Short Answer

Error 0B83: Inconsistency between mA anode and mA cathode measurement.

Possible Causes

Subsystem/category: mA. Manual recovery/error class: ABORT.

Troubleshooting Steps

Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Error 0B84

Error Code System Message

​

0B84, error 0B84, XR240 0B84, Optima XR240amx error 0B84, mA measurement at start of exposure lower than expected
Question

What does error 0B84 mean on a GE Optima XR240amx?

Short Answer

Error 0B84: mA measurement at start of exposure lower than expected.

Possible Causes

Subsystem/category: mA. Manual recovery/error class: DEBUG.

Troubleshooting Steps

Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Error 0B85

Error Code System Message

​

0B85, error 0B85, XR240 0B85, Optima XR240amx error 0B85, mA measurement at start of exposure higher than expected
Question

What does error 0B85 mean on a GE Optima XR240amx?

Short Answer

Error 0B85: mA measurement at start of exposure higher than expected.

Possible Causes

Subsystem/category: mA. Manual recovery/error class: DEBUG.

Troubleshooting Steps

Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Error 0B86

Error Code System Message

​

0B86, error 0B86, XR240 0B86, Optima XR240amx error 0B86, Filament aging saturation detected when computed target reaches command limit
Question

What does error 0B86 mean on a GE Optima XR240amx?

Short Answer

Error 0B86: Filament aging saturation detected when computed target reaches command limit.

Possible Causes

Subsystem/category: mA. Manual recovery/error class: DEBUG.

Troubleshooting Steps

Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts.

Parts Areas to Check

​

For Related Models

Optima XR240amx