AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 0B82
Error Code System Message
Question
What does error 0B82 mean on a GE Optima XR240amx?
Short Answer
Error 0B82: No mAs measured after exposure start.
Possible Causes
Subsystem/category: mA. Manual recovery/error class: ABORT.
Troubleshooting Steps
Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0B83
Error Code System Message
Question
What does error 0B83 mean on a GE Optima XR240amx?
Short Answer
Error 0B83: Inconsistency between mA anode and mA cathode measurement.
Possible Causes
Subsystem/category: mA. Manual recovery/error class: ABORT.
Troubleshooting Steps
Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0B84
Error Code System Message
Question
What does error 0B84 mean on a GE Optima XR240amx?
Short Answer
Error 0B84: mA measurement at start of exposure lower than expected.
Possible Causes
Subsystem/category: mA. Manual recovery/error class: DEBUG.
Troubleshooting Steps
Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0B85
Error Code System Message
Question
What does error 0B85 mean on a GE Optima XR240amx?
Short Answer
Error 0B85: mA measurement at start of exposure higher than expected.
Possible Causes
Subsystem/category: mA. Manual recovery/error class: DEBUG.
Troubleshooting Steps
Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0B86
Error Code System Message
Question
What does error 0B86 mean on a GE Optima XR240amx?
Short Answer
Error 0B86: Filament aging saturation detected when computed target reaches command limit.
Possible Causes
Subsystem/category: mA. Manual recovery/error class: DEBUG.
Troubleshooting Steps
Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts.
Parts Areas to Check