AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 1221102
Error Code System Message
Question
What does error 1221102 mean on a GE Optima XR240amx?
Short Answer
Error 1221102: Detector calibration upload completed.
Possible Causes
Subsystem/category: Detector. Manual recovery/error class: WARNING.
Troubleshooting Steps
This is a successful status message; proceed with registration/use as appropriate.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 1221217
Error Code System Message
Question
What does error 1231217 mean on a GE Optima XR240amx?
Short Answer
Error 1231217: A non-compatible detector has been connected to the system.
Possible Causes
Subsystem/category: Detector. Manual recovery/error class: WARNING.
Troubleshooting Steps
Remove the non-compatible detector. Verify detector vintage, private-key applicability and compatibility before attempting registration again.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 1221220
Error Code System Message
Question
What does error 1221220 mean on a GE Optima XR240amx?
Short Answer
Error 1221220: Digital Cassette detector data upload failed.
Possible Causes
Subsystem/category: Detector. Manual recovery/error class: WARNING.
Troubleshooting Steps
Reboot the system and retry. If the problem persists, troubleshoot detector communication and escalate to service.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 1221221
Error Code System Message
Question
What does error 1221221 mean on a GE Optima XR240amx?
Short Answer
Error 1221221: Digital cassette detector is not compatible with the system.
Possible Causes
Subsystem/category: Detector. Manual recovery/error class: WARNING.
Troubleshooting Steps
Verify detector vintage, firmware, SDK and compatibility table; update only through the approved detector firmware/compatibility procedure.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 1221232
Error Code System Message
Question
What does error 1221232 mean on a GE Optima XR240amx?
Short Answer
Error 1221232: Detector data upload was interrupted.
Possible Causes
Subsystem/category: Detector. Manual recovery/error class: WARNING.
Troubleshooting Steps
Verify detector communication, reboot/retry as appropriate, and isolate tether versus wireless communication if the interruption repeats.
Parts Areas to Check