AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 1221243
Error Code System Message
Question
What does error 1221243 mean on a GE Optima XR240amx?
Short Answer
Error 1221243: Wireless access point communication failed.
Possible Causes
Subsystem/category: Wireless AP. Manual recovery/error class: WARNING.
Troubleshooting Steps
Test AP communication, verify power/Ethernet and country configuration, then reboot. If corruption is suspected, use the approved AP reset/reconfiguration procedure.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 1221244
Error Code System Message
Question
What does error 1221244 mean on a GE Optima XR240amx?
Short Answer
Error 1221244: Access point configuration failed.
Possible Causes
Subsystem/category: Wireless AP. Manual recovery/error class: WARNING.
Troubleshooting Steps
Re-select/commit the correct country code. If failure persists, verify AP firmware and whether a permitted detector/AP channel intersection exists.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 1221245
Error Code System Message
Question
What does error 1221245 mean on a GE Optima XR240amx?
Short Answer
Error 1221245: Access point boot failed.
Possible Causes
Subsystem/category: Wireless AP. Manual recovery/error class: WARNING.
Troubleshooting Steps
Wait for boot attempt to finish, reboot the system, then verify country configuration, AP Ethernet/power connections and AP operation.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 3600124
Error Code System Message
Question
What does error 3600124 mean on a GE Optima XR240amx?
Short Answer
Error 3600124: Generator Communication Failure.
Possible Causes
Subsystem/category: Generator. Manual recovery/error class: ERROR.
Troubleshooting Steps
Review generator error history, power and communication status, related Djinn errors and cabling before ordering parts.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 3606024
Error Code System Message
Question
What does error 3606024 mean on a GE Optima XR240amx?
Short Answer
Error 3606024: Generator Power Up Failure.
Possible Causes
Subsystem/category: Generator. Manual recovery/error class: ERROR.
Troubleshooting Steps
Review the Class M error log and related generator errors, then verify generator power, communications and wiring before parts replacement.
Parts Areas to Check