AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 321

Error Code System Message

321

AMX-4, AMX-4+, Definium AMX 700, ERROR 321, Error 321
Question

What should I check for error 321 on the AMX-4 Series?

Short Answer

1) Check Items 1 and 3 from Error 320 2) Check for 27 VAC at J4-6 & 7 on Rotor Control Board 3) Check for 60 Hz Inverter signal OK signal at J1-5 on the Rotor Control Board 4) If Error 320 occurs after Error 321, there may be a temperature problem on the 60 Hz Inverter

Possible Causes

1) Check Items 1 and 3 from Error 320 2) Check for 27 VAC at J4-6 & 7 on Rotor Control Board 3) Check for 60 Hz Inverter signal OK signal at J1-5 on the Rotor Control Board 4) If Error 320 occurs after Error 321, there may be a temperature problem on the 60 Hz Inverter

Troubleshooting Steps

1) Check Items 1 and 3 from Error 320 2) Check for 27 VAC at J4-6 & 7 on Rotor Control Board 3) Check for 60 Hz Inverter signal OK signal at J1-5 on the Rotor Control Board 4) If Error 320 occurs after Error 321, there may be a temperature problem on the 60 Hz Inverter

Parts Areas to Check

Rotor Control Board; Rotor

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 322

Error Code System Message

322

AMX-4, AMX-4+, Definium AMX 700, ERROR 322, Error 322
Question

What should I check for error 322 on the AMX-4 Series?

Short Answer

1) Check Battery Charger Board feedback at U123(U77) Pin 5 of CPU Board 2) Check Battery Charger Board feedback and make sure it is going out to the CPU (TP-6 on the Battery Charger Board) 3) Check the ribbon cable from the Battery Charger Board to the CPU Board 4) Batteries may be defective If Voltage Is Really Less than 90 volts 5) Bad Battery Charger Board 6) Loose Battery terminals

Possible Causes

1) Check Battery Charger Board feedback at U123(U77) Pin 5 of CPU Board 2) Check Battery Charger Board feedback and make sure it is going out to the CPU (TP-6 on the Battery Charger Board) 3) Check the ribbon cable from the Battery Charger Board to the CPU Board 4) Batteries may be defective If Voltage Is Really Less than 90 volts 5) Bad Battery Charger Board 6) Loose Battery terminals. 7) Check for possible bad LVLE Power Supply 8) Check for possible bad 1kHz board 9) If Rotor Controller Board oveheat, circuit breaker tripped tooor same time

Troubleshooting Steps

1) Check Battery Charger Board feedback at U123(U77) Pin 5 of CPU Board 2) Check Battery Charger Board feedback and make sure it is going out to the CPU (TP-6 on the Battery Charger Board) 3) Check the ribbon cable from the Battery Charger Board to the CPU Board 4) Batteries may be defective If Voltage Is Really Less than 90 volts 5) Bad Battery Charger Board 6) Loose Battery terminals

Parts Areas to Check

CPU Board; Battery Charger Board; Battery

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 322

Error Code System Message

ERROR 322 - Battery voltage very low

ERROR 322, Battery voltage very low, AMX-4 ERROR 322, AMX-4+ ERROR 322, Definium AMX 700 ERROR 322, GE AMX error ERROR 322
Question

What does ERROR 322 mean on a GE AMX-Series mobile X-ray?

Short Answer

ERROR 322: Battery voltage very low.

Possible Causes

Battery voltage was below the expected minimum.

Troubleshooting Steps

Check actual battery voltage, CPU/charger feedback and battery-voltage calibration.

Parts Areas to Check

Battery set; charger board; CPU board; voltage/current feedback wiring

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 326

Error Code System Message

326

AMX-4, AMX-4+, Definium AMX 700, ERROR 326, Error 326
Question

What should I check for error 326 on the AMX-4 Series?

Short Answer

1) Check items 1, 2, and 3 from Error 322 2) Batteries may be defective If Voltage Is Really More Than 150 volts Note: If unit does not charge (no voltage across sense resistor AMX1 A3 R1) - verify that the safety contactor contacts are not welded shut

Possible Causes

1) Check items 1, 2, and 3 from Error 322 2) Batteries may be defective If Voltage Is Really More Than 150 volts Note: If unit does not charge (no voltage across sense resistor AMX1 A3 R1) - verify that the safety contactor contacts are not welded shut

Troubleshooting Steps

1) Check items 1, 2, and 3 from Error 322 2) Batteries may be defective If Voltage Is Really More Than 150 volts Note: If unit does not charge (no voltage across sense resistor AMX1 A3 R1) - verify that the safety contactor contacts are not welded shut

Parts Areas to Check

Safety Contactor

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 326

Error Code System Message

ERROR 326 - Battery voltage above expected range

ERROR 326, Battery voltage above expected range, AMX-4 ERROR 326, AMX-4+ ERROR 326, Definium AMX 700 ERROR 326, GE AMX error ERROR 326
Question

What does ERROR 326 mean on a GE AMX-Series mobile X-ray?

Short Answer

ERROR 326: Battery voltage above expected range.

Possible Causes

Battery voltage exceeded expected upper range.

Troubleshooting Steps

Verify battery voltage and calibration; inspect CPU/charger feedback path.

Parts Areas to Check

Battery set; charger board; CPU board; voltage/current feedback wiring

For Related Models

AMX-4, AMX-4+, Definium AMX 700