AMX Troubleshooting & Technical Support
Results Count
Error Number
CAL CHGR ERR 1
Error Code System Message
CAL CHGR ERR 1 - Charger calibration counter overflow
Question
What does CAL CHGR ERR 1 mean on a GE AMX-Series mobile X-ray?
Short Answer
CAL CHGR ERR 1: Charger calibration counter overflow.
Possible Causes
Charge-current calibration count overflowed.
Troubleshooting Steps
Check charger-board frequency/current feedback and CPU connection.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
CAL CHGR ERR 2
Error Code System Message
CAL CHGR ERR 2 - Charge-current polarity/range invalid
Question
What does CAL CHGR ERR 2 mean on a GE AMX-Series mobile X-ray?
Short Answer
CAL CHGR ERR 2: Charge-current polarity/range invalid.
Possible Causes
Floating charge-current input indicated an unexpected value during calibration.
Troubleshooting Steps
Check CPU-to-charger connection and charger-current calibration circuit.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
CAL CHGR ERR 3
Error Code System Message
CAL CHGR ERR 3 - Charger calibration limit exceeded
Question
What does CAL CHGR ERR 3 mean on a GE AMX-Series mobile X-ray?
Short Answer
CAL CHGR ERR 3: Charger calibration limit exceeded.
Possible Causes
Upper/lower charger calibration count limit exceeded.
Troubleshooting Steps
Calibrate charger; inspect charger-board/CPU feedback and charge-scale-select path.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
CAL DAT WARNG
Error Code System Message
CAL DAT WARNG - Calibration-data warning
Question
What does CAL DAT WARNG mean on a GE AMX-Series mobile X-ray?
Short Answer
CAL DAT WARNG: Calibration-data warning.
Possible Causes
A calibration parameter is out of limit; some calibration may continue with warning.
Troubleshooting Steps
Review the affected calibration values and correct the out-of-range parameter before finalizing.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
CAL TAP ERROR 1
Error Code System Message
CAL TAP ERROR 1 - Proper generator mA could not be reached
Question
What does CAL TAP ERROR 1 mean on a GE AMX-Series mobile X-ray?
Short Answer
CAL TAP ERROR 1: Proper generator mA could not be reached.
Possible Causes
Changing the kVp/tap could not achieve the required mA.
Troubleshooting Steps
Check mA calibration, load connections and repeat tap calibration.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error