AMX Troubleshooting & Technical Support

Results Count

Error Number

TEST 03 FAILED

Error Code System Message

TEST 03 FAILED - Auxiliary/external RAM test failed

TEST 03 FAILED, Auxiliary/external RAM test failed, AMX-4 TEST 03 FAILED, AMX-4+ TEST 03 FAILED, Definium AMX 700 TEST 03 FAILED, GE AMX error TEST 03 FAILED
Question

What does TEST 03 FAILED mean on a GE AMX-Series mobile X-ray?

Short Answer

TEST 03 FAILED: Auxiliary/external RAM test failed.

Possible Causes

External RAM test patterns could not be written/read correctly. Preserve data form bottom of RAM to top of Data Base. Does not preserve data abovetop of Data Base. Error Code 83 Hex is recorded in the Error Log for a failure.

Troubleshooting Steps

Check RAM devices and associated circuitry. Check RMA and it's associated circuity. Make sure RAMis in the correct socket. Check the microprocessor. Check the option switch. S96 on CPU 46-232828 or S75 on CPU 46-264974, to see that RAM is selected. Option switch 1 must be in the On (closed) position. Closed = 2k of RAM. Open = 4k of RAM. Change RAM chip first on CPU board U54 abd U107. If it still not work, the LVLE Power Supply is bad and need to be replace.

Parts Areas to Check

CPU board; memory/timer/A-D circuitry identified by the test

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

TEST 04 FAILED

Error Code System Message

TEST 04 FAILED - Calibration data checksum failed

TEST 04 FAILED, Calibration data checksum failed, AMX-4 TEST 04 FAILED, AMX-4+ TEST 04 FAILED, Definium AMX 700 TEST 04 FAILED, GE AMX error TEST 04 FAILED
Question

What does TEST 04 FAILED mean on a GE AMX-Series mobile X-ray?

Short Answer

TEST 04 FAILED: Calibration data checksum failed.

Possible Causes

One or more calibration-data records do not match their stored checksum. Calculates the Check Sum for the Calibration Data Base and compares it with the stored value. Test fails if they are not equal. Entering Calibration will give a more specific error code for the check sum failure. Error code 84 Hex is recorded in the Error Code for a failure.

Troubleshooting Steps

Check Calibration and RAM. Identify the affected calibration, correct it, and rerun checksum/calibration.

Parts Areas to Check

CPU board; memory/timer/A-D circuitry identified by the test

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

TEST 05 FAILED

Error Code System Message

TEST 05 FAILED - Watchdog timer test failed

TEST 05 FAILED, Watchdog timer test failed, AMX-4 TEST 05 FAILED, AMX-4+ TEST 05 FAILED, Definium AMX 700 TEST 05 FAILED, GE AMX error TEST 05 FAILED
Question

What does TEST 05 FAILED mean on a GE AMX-Series mobile X-ray?

Short Answer

TEST 05 FAILED: Watchdog timer test failed.

Possible Causes

Write to timer and read back several times. Test fails if timer is high for more than 75 milliseconds or less than 25 milliseconds. Error Code 85 Hex is recorded in the Error Log for a failure. Error Watchdog timing/reset response is outside expected limits.

Troubleshooting Steps

Check HC123 on shot and associated circuitry. Check watchdog timer circuitry and CPU timing components.

Parts Areas to Check

CPU board; memory/timer/A-D circuitry identified by the test

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

TEST 06 FAILED

Error Code System Message

TEST 06 FAILED - Programmable interval timer test failed

TEST 06 FAILED, Programmable interval timer test failed, AMX-4 TEST 06 FAILED, AMX-4+ TEST 06 FAILED, Definium AMX 700 TEST 06 FAILED, GE AMX error TEST 06 FAILED
Question

What does TEST 06 FAILED mean on a GE AMX-Series mobile X-ray?

Short Answer

TEST 06 FAILED: Programmable interval timer test failed.

Possible Causes

Timer 1 cannot be checked because the program does not have direct control of it's input gate. Initialize timer 0 for 16 kHz and timer 2 for 120 kHz, Count pulses to verify frequency. Check output 2 for 50% duty cycle. Error Code 85 Hex is recorded in the Error Log for a failure. Timer could not generate the required interval/count behavior.

Troubleshooting Steps

Check 82C54 timer, HC151 8 to 1 MUX, oscillator, and all associated circuitry. Check interval-timer IC/clock and associated CPU circuitry.

Parts Areas to Check

CPU board; memory/timer/A-D circuitry identified by the test

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

TEST 07 FAILED

Error Code System Message

TEST 07 FAILED - Analog-to-digital converter test failed

TEST 07 FAILED, Analog-to-digital converter test failed, AMX-4 TEST 07 FAILED, AMX-4+ TEST 07 FAILED, Definium AMX 700 TEST 07 FAILED, GE AMX error TEST 07 FAILED
Question

What does TEST 07 FAILED mean on a GE AMX-Series mobile X-ray?

Short Answer

TEST 07 FAILED: Analog-to-digital converter test failed.

Possible Causes

Read +5 volt power supply. Test fails if result is below +4.5 volts or above +5.5 volts. Error Code 87 Hex is recorded in the Error Log for a failure. Also test +24V and +/- 15 supplies if option switch is set to do so. (+24V and +/- 15V supply test only with firmware 46-302688G1 / 46-302687G1 and later on CPU 46-264974. Test enabled by dip switch #3.) A/D converter reference/power-supply input is outside expected range.

Troubleshooting Steps

Check +5 volt supply. AD574A A/D Converter LF398 sample abd hold, AD5706 16 to 1 MUX, HC374 latch, and all associated circuitry. Check +5 V/reference supply, A/D converter and CPU board circuitry. If display showed "DRIVE WORD LIMIT" then RAM chips is bad, maybe bad U342 and U413.

Parts Areas to Check

CPU board; memory/timer/A-D circuitry identified by the test

For Related Models

AMX-4, AMX-4+, Definium AMX 700