AMX Troubleshooting & Technical Support

Results Count

Error Number

X-RAY WORD LIMIT

Error Code System Message

X-RAY WORD LIMIT

AMX X-RAY WORD LIMIT, CAL FIL CUR TBL
Question

How do I recover from X-RAY WORD LIMIT during AMX-Series calibration?

Short Answer

The manual describes resetting the affected calibration state and re-running the relevant mAs/current calibration before repeating CAL FIL CUR TBL.

Possible Causes

Calibration parameter exceeded allowable word range.

Troubleshooting Steps

Follow the manual's service reset sequence for the affected calibration data, redo the required mAs/current calibration, then rerun the filament-current-table calibration.

Parts Areas to Check

CPU calibration data; generator calibration

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

X-RAY WORD LIMIT

Error Code System Message

X-RAY WORD LIMIT - X-ray calibration word limit exceeded

X-RAY WORD LIMIT, X-ray calibration word limit exceeded, AMX-4 X-RAY WORD LIMIT, AMX-4+ X-RAY WORD LIMIT, Definium AMX 700 X-RAY WORD LIMIT, GE AMX error X-RAY WORD LIMIT
Question

What does X-RAY WORD LIMIT mean on a GE AMX-Series mobile X-ray?

Short Answer

X-RAY WORD LIMIT: X-ray calibration word limit exceeded.

Possible Causes

One or more X-ray calibration parameters exceeded word range.

Troubleshooting Steps

Follow the generator-calibration recovery procedure and repeat the affected calibration.

Parts Areas to Check

CPU board; calibration data; subsystem identified by the error

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

X-Ray Control Module smoking, no errors, kVp waveform OK

Error Code System Message

​

AMX-4, AMX-4+, Definium AMX 700, AMCI-AMX4-TS-0154, AMX1A4R7 smoking - no errors - kVp Waveform OK
Question

How do I troubleshoot this AMX-4 Series issue: AMX1A4R7 smoking - no errors - kVp Waveform OK?

Short Answer

Replace bad 1 KHZ Inverter Driver board (Primary Current P1/P2 Approx. 80 Amps Vice 50-60 Amps) – Also Replace X-Ray Control Module

Possible Causes

Problem: AMX1A4R7 smoking - no errors - kVp Waveform OK

Troubleshooting Steps

Replace bad 1 KHZ Inverter Driver board (Primary Current P1/P2 Approx. 80 Amps Vice 50-60 Amps) – Also Replace AMX1A4R7

Parts Areas to Check

1 KHZ Inverter Driver Board

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

X-Ray On Light” Or Sonalert Comes On During Charge Mode Or During Driving Or During Idle Mode

Error Code System Message

​

AMX-4, AMX-4+, Definium AMX 700, AMCI-AMX4-TS-0162, Intermittent “X-Ray On Light” Or Sonalert Comes On During Charge Mode Or During Driving Or During Idle Mode
Question

How do I troubleshoot this AMX-4 Series issue: Intermittent “X-Ray On Light” Or Sonalert Comes On During Charge Mode Or During Driving Or During Idle Mode?

Short Answer

Replace The LVLE Power Supply (T3497L)

Possible Causes

Problem: Intermittent “X-Ray On Light” Or Sonalert Comes On During Charge Mode Or During Driving Or During Idle Mode

Troubleshooting Steps

Replace The LVLE Power Supply (T3497L)

Parts Areas to Check

LVLE Power Supply

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

X-ray terminates early

Error Code System Message

ERROR 462 / 466 / 467 / 468

AMX exposure stops early, long exposure, hardware backup timer
Question

What can cause an AMX-Series exposure to terminate early?

Short Answer

The manual lists excessive exposure time, hardware backup timer expiration, explicit exposure-stop state, interlock changes, or generator feedback faults as possible causes.

Possible Causes

Timing limit reached; stop command; interlock opened; generator output/feedback mismatch.

Troubleshooting Steps

Review ERROR 462, 466, 467, 468 and related generator codes. Check exposure time, interlocks, timing backup circuits, command/feedback status, and battery voltage under load.

Parts Areas to Check

Generator timing; CPU; interlocks; battery condition

For Related Models

AMX-4, AMX-4+, Definium AMX 700