AMX Troubleshooting & Technical Support

Results Count

Error Number

mA appears High and kVp low-Can’t get the mA down

Error Code System Message

​

AMX-4, AMX-4+, Definium AMX 700, AMCI-AMX4-TS-0153, mA appears High and kVp low-Can’t get the mA down
Question

How do I troubleshoot this AMX-4 Series issue: mA appears High and kVp low-Can’t get the mA down?

Short Answer

Check To See That The Small Spot Is Boosting Rather Than The Large Spot (Filament Mis-wire In The XRT)

Possible Causes

Problem: mA appears High and kVp low-Can’t get the mA down

Troubleshooting Steps

Check To See That The Small Spot Is Boosting Rather Than The Large Spot (Filament Mis-wire In The XRT)

Parts Areas to Check

​

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

mAs calibration error

Error Code System Message

MAS CAL ERROR 1 / 4

AMX mAs calibration, MAS CAL ERROR
Question

How does AMX-Series mAs calibration work?

Short Answer

The system injects a known current into the mA measuring circuit and compares counted pulses to expected mAs response.

Possible Causes

mAs VCO/counter/calibration fault.

Troubleshooting Steps

Verify the calibration source/current, mA measuring circuit, VCO/counter response and CPU data before repeating calibration.

Parts Areas to Check

mA measurement circuit; VCO; CPU counter

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

system is switched from an “OFF” state to “ON” state

Error Code System Message

​

AMX-4, AMX-4+, Definium AMX 700, AMCI-AMX4-TS-0176, When the AMX4/AMX4+ system is switched from an “OFF” state to “ON” state, a peak inrush current spike greater than the 10A rating of the K215 relay contacts was observed. This spike has the potential to weld the contacts
Question

How do I troubleshoot this AMX-4 Series issue: When the AMX4/AMX4+ system is switched from an “OFF” state to “ON” state, a peak inrush current spike greater than the 10A rating of the K215 relay contacts was observed. This spike has the potential to weld the contacts?

Short Answer

The solution is to replace the CONDOR power supply (Vendor part# MDSP1733, GE part# 2318476) with the modified CONDOR/SL power supply (Vendor part# MDSP1733A, Orderable GE part# 2366118). The new supply has an inrush current profile that is much smaller than the existing supply. Refer to Service Note 30004.

Possible Causes

Problem: When the AMX4/AMX4+ system is switched from an “OFF” state to “ON” state, a peak inrush current spike greater than the 10A rating of the K215 relay contacts was observed. This spike has the potential to weld the contacts of the K215 relay. This welding action causes the system to remain in the “ON “state even after switching the state to “OFF.”

Troubleshooting Steps

The solution is to replace the CONDOR power supply (Vendor part# MDSP1733, GE part# 2318476) with the modified CONDOR/SL power supply (Vendor part# MDSP1733A, Orderable GE part# 2366118). The new supply has an inrush current profile that is much smaller than the existing supply. Refer to Service Note 30004.

Parts Areas to Check

​

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

AC not present

Error Code System Message

​

Brivo XR285 AC present not detected
Question

What does the AC Present diagnostic test on the Brivo XR285amx?

Short Answer

It verifies that the system recognizes when the AC power plug is connected to the receptacle.

Possible Causes

AC-detect circuit issue; plug/receptacle problem; power-control communication fault.

Troubleshooting Steps

Connect AC power and run AC Present. If the system does not recognize AC, verify the physical connection and then troubleshoot the power-control detection path.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

BDM mismatch

Error Code System Message

​

Brivo XR285amx BDM mismatch restore cancel
Question

What should I do when the Brivo XR285amx shows a BDM Mismatch screen?

Short Answer

For a normally operating system with no parts replaced, select RESTORE. If parts were replaced, follow the replacement-finalization instructions for that component.

Possible Causes

Configuration fingerprint or firmware mismatch detected during power-up.

Troubleshooting Steps

Confirm whether any intelligent board or generator component was replaced. If no parts were changed, select RESTORE. If a part was replaced, use the System Manual finalization procedure for the correct Restore/Cancel selection.

Parts Areas to Check

​

For Related Models

Brivo XR285amx