top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 0x0183

Error Code System Message

​

0x0183, 0X0183, Brivo XR285amx 0x0183, Djinn Heater 1 error, Id SW Filament 1 driver diag failed
Question

What does 0x0183 mean on a GE Brivo XR285amx?

Short Answer

0x0183: Id SW Filament 1 driver diag failed.

Possible Causes

The Djinn Beetle generator reported a Heater 1 SW fault: Id SW Filament 1 driver diag failed.

Troubleshooting Steps

Run Fil 1 Drv Clsd Loop, Fil 1 Xfmr Primary, and Fil 1 Drv Short Test as applicable. Check the J5/J6 filament power connection, cathode HV cable, filament impedance, and 48 V supply before isolating an FRU.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0200

Error Code System Message

​

0x0200, 0X0200, Brivo XR285amx 0x0200, Djinn Heater 2 error, Unkwn FW error
Question

What does 0x0200 mean on a GE Brivo XR285amx?

Short Answer

0x0200: Unkwn FW error.

Possible Causes

The Djinn Beetle generator reported a Heater 2 FW fault: Unkwn FW error.

Troubleshooting Steps

Run Fil 2 Drv Clsd Loop, Fil 2 Xfmr Primary, and Fil 2 Drv Short Test as applicable. Check the J5/J6 filament power connection, cathode HV cable, filament impedance, and 48 V supply before isolating an FRU.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0201

Error Code System Message

​

0x0201, 0X0201, Brivo XR285amx 0x0201, Djinn Heater 2 error, Soft over current (measured current > estimated current)
Question

What does 0x0201 mean on a GE Brivo XR285amx?

Short Answer

0x0201: Soft over current (measured current > estimated current).

Possible Causes

The Djinn Beetle generator reported a Heater 2 FW fault: Soft over current (measured current > estimated current).

Troubleshooting Steps

Run Fil 2 Drv Clsd Loop, Fil 2 Xfmr Primary, and Fil 2 Drv Short Test as applicable. Check the J5/J6 filament power connection, cathode HV cable, filament impedance, and 48 V supply before isolating an FRU.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0202

Error Code System Message

​

0x0202, 0X0202, Brivo XR285amx 0x0202, Djinn Heater 2 error, Soft open circuit (measured current
Question

What does 0x0202 mean on a GE Brivo XR285amx?

Short Answer

0x0202: Soft open circuit (measured current < estimated current).

Possible Causes

The Djinn Beetle generator reported a Heater 2 FW fault: Soft open circuit (measured current < estimated current).

Troubleshooting Steps

Run Fil 2 Drv Clsd Loop, Fil 2 Xfmr Primary, and Fil 2 Drv Short Test as applicable. Check the J5/J6 filament power connection, cathode HV cable, filament impedance, and 48 V supply before isolating an FRU.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0203

Error Code System Message

​

0x0203, 0X0203, Brivo XR285amx 0x0203, Djinn Heater 2 error, Demand too high (target > max heater)
Question

What does 0x0203 mean on a GE Brivo XR285amx?

Short Answer

0x0203: Demand too high (target > max heater).

Possible Causes

The Djinn Beetle generator reported a Heater 2 FW fault: Demand too high (target > max heater).

Troubleshooting Steps

Run Fil 2 Drv Clsd Loop, Fil 2 Xfmr Primary, and Fil 2 Drv Short Test as applicable. Check the J5/J6 filament power connection, cathode HV cable, filament impedance, and 48 V supply before isolating an FRU.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

bottom of page