AMX Troubleshooting & Technical Support
Results Count
Error Number
CAL TAP ERROR 7
Error Code System Message
CAL TAP ERROR 7 - Tap calibration parameter limit exceeded
Question
What does CAL TAP ERROR 7 mean on a GE AMX-Series mobile X-ray?
Short Answer
CAL TAP ERROR 7: Tap calibration parameter limit exceeded.
Possible Causes
One or more tap calibration parameters exceeded limits.
Troubleshooting Steps
Repeat Tap Calibration and inspect generator/tap components.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
CAL TUBE ERR 1
Error Code System Message
CAL TUBE ERR 1 - Tube-calibration parameter limit exceeded
Question
What does CAL TUBE ERR 1 mean on a GE AMX-Series mobile X-ray?
Short Answer
CAL TUBE ERR 1: Tube-calibration parameter limit exceeded.
Possible Causes
One or more tube calibration parameters exceeded allowed limits.
Troubleshooting Steps
Refer to the specific limit/error and correct tube/generator calibration.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
CHARG BYTE LIMIT
Error Code System Message
CHARG BYTE LIMIT - Charger byte calibration limit exceeded
Question
What does CHARG BYTE LIMIT mean on a GE AMX-Series mobile X-ray?
Short Answer
CHARG BYTE LIMIT: Charger byte calibration limit exceeded.
Possible Causes
One or more charger calibration values exceeded byte limits.
Troubleshooting Steps
Calibrate charger and inspect charger-feedback circuitry.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
CHARG WORD LIMIT
Error Code System Message
CHARG WORD LIMIT - Charger word calibration limit exceeded
Question
What does CHARG WORD LIMIT mean on a GE AMX-Series mobile X-ray?
Short Answer
CHARG WORD LIMIT: Charger word calibration limit exceeded.
Possible Causes
Charge-current calibration parameter exceeded word range.
Troubleshooting Steps
Calibrate charger; check CPU/charger-board connections and charge-scale-select logic.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
CHECK SUM ERR 1
Error Code System Message
CHECK SUM ERR 1 - Drive calibration checksum failed
Question
What does CHECK SUM ERR 1 mean on a GE AMX-Series mobile X-ray?
Short Answer
CHECK SUM ERR 1: Drive calibration checksum failed.
Possible Causes
Drive calibration data checksum failed.
Troubleshooting Steps
Recalibrate drive.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error