AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 0x0582

Error Code System Message

​

0x0582, 0X0582, Brivo XR285amx 0x0582, Djinn BB error, input voltage detected too high
Question

What does 0x0582 mean on a GE Brivo XR285amx?

Short Answer

0x0582: input voltage detected too high.

Possible Causes

The Djinn Beetle generator reported a BB SW fault: input voltage detected too high.

Troubleshooting Steps

Run Rise Time/Regulation, Power Supplies, and relevant DC-bus checks. Verify battery input is at least 144 VDC, inspect the DC-bus cable and breaker path, and confirm the 48 V supply is within specification.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0583

Error Code System Message

​

0x0583, 0X0583, Brivo XR285amx 0x0583, Djinn BB error, DC bus voltage too low
Question

What does 0x0583 mean on a GE Brivo XR285amx?

Short Answer

0x0583: DC bus voltage too low.

Possible Causes

The Djinn Beetle generator reported a BB SW fault: DC bus voltage too low.

Troubleshooting Steps

Run Rise Time/Regulation, Power Supplies, and relevant DC-bus checks. Verify battery input is at least 144 VDC, inspect the DC-bus cable and breaker path, and confirm the 48 V supply is within specification.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0584

Error Code System Message

​

0x0584, 0X0584, Brivo XR285amx 0x0584, Djinn BB error, input voltage detected
Question

What does 0x0584 mean on a GE Brivo XR285amx?

Short Answer

0x0584: input voltage detected.

Possible Causes

The Djinn Beetle generator reported a BB SW fault: input voltage detected.

Troubleshooting Steps

Run Rise Time/Regulation, Power Supplies, and relevant DC-bus checks. Verify battery input is at least 144 VDC, inspect the DC-bus cable and breaker path, and confirm the 48 V supply is within specification.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0585

Error Code System Message

​

0x0585, 0X0585, Brivo XR285amx 0x0585, Djinn BB error, Input current detected while BB is OFF
Question

What does 0x0585 mean on a GE Brivo XR285amx?

Short Answer

0x0585: Input current detected while BB is OFF.

Possible Causes

The Djinn Beetle generator reported a BB SW fault: Input current detected while BB is OFF.

Troubleshooting Steps

Run Rise Time/Regulation, Power Supplies, and relevant DC-bus checks. Verify battery input is at least 144 VDC, inspect the DC-bus cable and breaker path, and confirm the 48 V supply is within specification.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0600

Error Code System Message

​

0x0600, 0X0600, Brivo XR285amx 0x0600, Djinn kV error, Unknown FW error
Question

What does 0x0600 mean on a GE Brivo XR285amx?

Short Answer

0x0600: Unknown FW error.

Possible Causes

The Djinn Beetle generator reported a kV FW fault: Unknown FW error.

Troubleshooting Steps

Run KV Measurement, KV Test, Inverter Test, and Resonant Components as applicable. Check the HV Tank-to-Power Module measurement cable, DC-bus error data, inverter/tank connections, and HV cables for arcing when indicated.

Parts Areas to Check

​

For Related Models

Brivo XR285amx