top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 0x0606

Error Code System Message

​

0x0606, 0X0606, Brivo XR285amx 0x0606, Djinn kV error, Too many spits
Question

What does 0x0606 mean on a GE Brivo XR285amx?

Short Answer

0x0606: Too many spits.

Possible Causes

The Djinn Beetle generator reported a kV FW fault: Too many spits.

Troubleshooting Steps

Run KV Measurement, KV Test, Inverter Test, and Resonant Components as applicable. Check the HV Tank-to-Power Module measurement cable, DC-bus error data, inverter/tank connections, and HV cables for arcing when indicated.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0606

Error Code System Message

​

Brivo XR285 0606 too many spits, anode cathode arc
Question

What does error 0x0606 mean on the Brivo XR285amx?

Short Answer

0x0606 means the generator detected too many high-voltage spits/arcs.

Possible Causes

HV cable/tube/tank arcing; damaged insulation; poor HV connection.

Troubleshooting Steps

Read the error data to identify the arc side: 1 = anode, 2 = cathode, 3 = both. Inspect the corresponding HV cable connections for cracks, damaged pins, arc tracks, seating, and grease.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0607

Error Code System Message

​

0x0607, 0X0607, Brivo XR285amx 0x0607, Djinn kV error, Inverter resonant current time out
Question

What does 0x0607 mean on a GE Brivo XR285amx?

Short Answer

0x0607: Inverter resonant current time out.

Possible Causes

The Djinn Beetle generator reported a kV FW fault: Inverter resonant current time out.

Troubleshooting Steps

Run KV Measurement, KV Test, Inverter Test, and Resonant Components as applicable. Check the HV Tank-to-Power Module measurement cable, DC-bus error data, inverter/tank connections, and HV cables for arcing when indicated.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0608

Error Code System Message

​

0x0608, 0X0608, Brivo XR285amx 0x0608, Djinn kV error, Inverter over current
Question

What does 0x0608 mean on a GE Brivo XR285amx?

Short Answer

0x0608: Inverter over current.

Possible Causes

The Djinn Beetle generator reported a kV FW fault: Inverter over current.

Troubleshooting Steps

Run KV Measurement, KV Test, Inverter Test, and Resonant Components as applicable. Check the HV Tank-to-Power Module measurement cable, DC-bus error data, inverter/tank connections, and HV cables for arcing when indicated.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0609

Error Code System Message

​

0x0609, 0X0609, Brivo XR285amx 0x0609, Djinn kV error, kV not ready
Question

What does 0x0609 mean on a GE Brivo XR285amx?

Short Answer

0x0609: kV not ready.

Possible Causes

The Djinn Beetle generator reported a kV FW fault: kV not ready.

Troubleshooting Steps

Run KV Measurement, KV Test, Inverter Test, and Resonant Components as applicable. Check the HV Tank-to-Power Module measurement cable, DC-bus error data, inverter/tank connections, and HV cables for arcing when indicated.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

bottom of page