Error Number
CHECK SUM ERR 10
Error Code System Message
CHECK SUM ERR 10 - X-ray circuit-resistance checksum failed
Question
What does CHECK SUM ERR 10 mean on a GE AMX-Series mobile X-ray?
Short Answer
CHECK SUM ERR 10: X-ray circuit-resistance checksum failed.
Possible Causes
X-ray circuit-resistance calibration checksum failed.
Troubleshooting Steps
Calibrate generator.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
CHECK SUM ERR 11
Error Code System Message
CHECK SUM ERR 11 - Filament-current table checksum failed
Question
What does CHECK SUM ERR 11 mean on a GE AMX-Series mobile X-ray?
Short Answer
CHECK SUM ERR 11: Filament-current table checksum failed.
Possible Causes
Filament-current table checksum failed or was not calibrated.
Troubleshooting Steps
Calibrate filament current; if other generator checksums are present, calibrate generator first.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
CHECK SUM ERR 2
Error Code System Message
CHECK SUM ERR 2 - Drive calibration word checksum failed
Question
What does CHECK SUM ERR 2 mean on a GE AMX-Series mobile X-ray?
Short Answer
CHECK SUM ERR 2: Drive calibration word checksum failed.
Possible Causes
Drive calibration words failed checksum.
Troubleshooting Steps
Recalibrate drive.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
CHECK SUM ERR 3
Error Code System Message
CHECK SUM ERR 3 - Charger calibration checksum failed
Question
What does CHECK SUM ERR 3 mean on a GE AMX-Series mobile X-ray?
Short Answer
CHECK SUM ERR 3: Charger calibration checksum failed.
Possible Causes
Charger calibration bytes failed checksum.
Troubleshooting Steps
Recalibrate charger.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
CHECK SUM ERR 4
Error Code System Message
CHECK SUM ERR 4 - Charger calibration word checksum failed
Question
What does CHECK SUM ERR 4 mean on a GE AMX-Series mobile X-ray?
Short Answer
CHECK SUM ERR 4: Charger calibration word checksum failed.
Possible Causes
Charger calibration words failed checksum.
Troubleshooting Steps
Recalibrate charger.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error