top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 0x0787

Error Code System Message

​

0x0787, 0X0787, Brivo XR285amx 0x0787, Djinn Thermal error, Single Wire communication impossible
Question

What does 0x0787 mean on a GE Brivo XR285amx?

Short Answer

0x0787: Single Wire communication impossible.

Possible Causes

The Djinn Beetle generator reported a Thermal SW fault: Single Wire communication impossible.

Troubleshooting Steps

Run Power Supply Thermal and the applicable thermal sensor tests. Check cooling, fans, sensor wiring, tube thermal-switch resistance, and single-wire communication before replacing a module.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0787

Error Code System Message

​

Brivo XR285 0787 single wire communication impossible
Question

What does error 0x0787 mean on the Brivo XR285amx?

Short Answer

0x0787: indicates single-wire communication is impossible.

Possible Causes

Single-wire thermal-device communication failure; tank/control connection issue.

Troubleshooting Steps

Use the Thermal Sensor no Tank test and related wiring checks. Verify thermal/ribbon cable connections and isolate whether the communication problem follows the tank or generator module.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0788

Error Code System Message

​

0x0788, 0X0788, Brivo XR285amx 0x0788, Djinn Thermal error, Filament-bias sensor failure
Question

What does 0x0788 mean on a GE Brivo XR285amx?

Short Answer

0x0788: Filament-bias sensor failure.

Possible Causes

The Djinn Beetle generator reported a Thermal SW fault: Filament-bias sensor failure.

Troubleshooting Steps

Run Power Supply Thermal and the applicable thermal sensor tests. Check cooling, fans, sensor wiring, tube thermal-switch resistance, and single-wire communication before replacing a module.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0789

Error Code System Message

​

0x0789, 0X0789, Brivo XR285amx 0x0789, Djinn Thermal error, Temperature conversion can’t be started
Question

What does 0x0789 mean on a GE Brivo XR285amx?

Short Answer

0x0789: Temperature conversion can’t be started.

Possible Causes

The Djinn Beetle generator reported a Thermal SW fault: Temperature conversion can’t be started.

Troubleshooting Steps

Run Power Supply Thermal and the applicable thermal sensor tests. Check cooling, fans, sensor wiring, tube thermal-switch resistance, and single-wire communication before replacing a module.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x078A

Error Code System Message

​

0x078A, 0X078A, Brivo XR285amx 0x078A, Djinn Thermal error, Thread controlling single wire devices is not responding
Question

What does 0x078A mean on a GE Brivo XR285amx?

Short Answer

0x078A: Thread controlling single wire devices is not responding.

Possible Causes

The Djinn Beetle generator reported a Thermal SW fault: Thread controlling single wire devices is not responding.

Troubleshooting Steps

Run Power Supply Thermal and the applicable thermal sensor tests. Check cooling, fans, sensor wiring, tube thermal-switch resistance, and single-wire communication before replacing a module.

Parts Areas to Check

​

For Related Models

Brivo XR285amx