AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 0x0902
Error Code System Message
Question
What does 0x0902 mean on a GE Brivo XR285amx?
Short Answer
0x0902: No software or software corrupted.
Possible Causes
The Djinn Beetle generator reported a Boot SW fault: No software or software corrupted.
Troubleshooting Steps
Review BDM/system self-recovery status and verify the Djinn software, database, FPGA, and rotation firmware are compatible. Load the correct 15 kW or 30 kW generator software/database when required.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0903
Error Code System Message
Question
What does 0x0903 mean on a GE Brivo XR285amx?
Short Answer
0x0903: No database or database corrupted.
Possible Causes
The Djinn Beetle generator reported a Boot SW fault: No database or database corrupted.
Troubleshooting Steps
Review BDM/system self-recovery status and verify the Djinn software, database, FPGA, and rotation firmware are compatible. Load the correct 15 kW or 30 kW generator software/database when required.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0905
Error Code System Message
Question
What does 0x0905 mean on a GE Brivo XR285amx?
Short Answer
0x0905: Database not compatible with software.
Possible Causes
The Djinn Beetle generator reported a Boot SW fault: Database not compatible with software.
Troubleshooting Steps
Review BDM/system self-recovery status and verify the Djinn software, database, FPGA, and rotation firmware are compatible. Load the correct 15 kW or 30 kW generator software/database when required.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0907
Error Code System Message
Question
What does 0x0907 mean on a GE Brivo XR285amx?
Short Answer
0x0907: FPGA missing or corrupted.
Possible Causes
The Djinn Beetle generator reported a Boot SW fault: FPGA missing or corrupted.
Troubleshooting Steps
Review BDM/system self-recovery status and verify the Djinn software, database, FPGA, and rotation firmware are compatible. Load the correct 15 kW or 30 kW generator software/database when required.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0908
Error Code System Message
Question
What does 0x0908 mean on a GE Brivo XR285amx?
Short Answer
0x0908: FPGA not compatible with application software.
Possible Causes
The Djinn Beetle generator reported a Boot SW fault: FPGA not compatible with application software.
Troubleshooting Steps
Review BDM/system self-recovery status and verify the Djinn software, database, FPGA, and rotation firmware are compatible. Load the correct 15 kW or 30 kW generator software/database when required.
Parts Areas to Check