AMX Troubleshooting & Technical Support
Results Count
Error Number
CHECK SUM ERR 5
Error Code System Message
CHECK SUM ERR 5 - Battery calibration checksum failed
Question
What does CHECK SUM ERR 5 mean on a GE AMX-Series mobile X-ray?
Short Answer
CHECK SUM ERR 5: Battery calibration checksum failed.
Possible Causes
Battery calibration words failed checksum.
Troubleshooting Steps
Calibrate voltmeter and generator as required.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
CHECK SUM ERR 6
Error Code System Message
CHECK SUM ERR 6 - X-ray calibration checksum failed
Question
What does CHECK SUM ERR 6 mean on a GE AMX-Series mobile X-ray?
Short Answer
CHECK SUM ERR 6: X-ray calibration checksum failed.
Possible Causes
X-ray calibration bytes failed checksum.
Troubleshooting Steps
Calibrate generator.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
CHECK SUM ERR 7
Error Code System Message
CHECK SUM ERR 7 - X-ray calibration word checksum failed
Question
What does CHECK SUM ERR 7 mean on a GE AMX-Series mobile X-ray?
Short Answer
CHECK SUM ERR 7: X-ray calibration word checksum failed.
Possible Causes
X-ray calibration words failed checksum.
Troubleshooting Steps
Calibrate generator.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
CHECK SUM ERR 8
Error Code System Message
CHECK SUM ERR 8 - Field-light calibration checksum failed
Question
What does CHECK SUM ERR 8 mean on a GE AMX-Series mobile X-ray?
Short Answer
CHECK SUM ERR 8: Field-light calibration checksum failed.
Possible Causes
Field-light calibration bytes failed checksum.
Troubleshooting Steps
Calibrate field light.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
CHECK SUM ERR 9
Error Code System Message
CHECK SUM ERR 9 - Turns-ratio calibration checksum failed
Question
What does CHECK SUM ERR 9 mean on a GE AMX-Series mobile X-ray?
Short Answer
CHECK SUM ERR 9: Turns-ratio calibration checksum failed.
Possible Causes
Turns-ratio data failed checksum.
Troubleshooting Steps
Calibrate generator.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error