top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Blinking RED - Detector warning

Error Code System Message

Blinking RED - Detector warning

Blinking RED, AMX 700 detector, Detector warning
Question

What does “Blinking RED” mean on a GE Definium AMX 700?

Short Answer

Blinking RED: Detector warning.

Possible Causes

Subsystem/category: Digital Detector. Source classification: LED STATUS.

Troubleshooting Steps

Proceed with caution and log the date/time and scenario.

Parts Areas to Check

TRAD digital detector

For Related Models

Definium AMX 700

Error Number

Both LEDs OFF - Detector has no power

Error Code System Message

Both LEDs OFF - Detector has no power

Both LEDs OFF, AMX 700 detector, Detector has no power
Question

What does “Both LEDs OFF” mean on a GE Definium AMX 700?

Short Answer

Both LEDs OFF: Detector has no power.

Possible Causes

Subsystem/category: Digital Detector. Source classification: LED STATUS.

Troubleshooting Steps

Close the exam and restart if the no-power state is unexpected.

Parts Areas to Check

TRAD digital detector

For Related Models

Definium AMX 700

Error Number

C-CODE HB

Error Code System Message

C-CODE HB - C-code heartbeat

C-CODE HB, DS1 GREEN blinking, AMX 700 C-CODE HB, C-code heartbeat
Question

What does “C-CODE HB” mean on a GE Definium AMX 700?

Short Answer

C-CODE HB: C-code heartbeat.

Possible Causes

Subsystem/category: Beetle Board. Source classification: LED STATUS.

Troubleshooting Steps

Interpret the LED together with the system symptom and use the matching Beetle/power/exposure troubleshooting path.

Parts Areas to Check

Beetle Interface Board

For Related Models

Definium AMX 700

Error Number

C-CODE HB LED

Error Code System Message

C-CODE HB - DS1 GREEN blinking

AMX 700 C-CODE HB LED, Beetle Board C-CODE HB, C-code heartbeat
Question

What does the C-CODE HB LED mean on the GE Definium AMX 700?

Short Answer

Blinking indicates embedded-processor C-code is running correctly.

Possible Causes

C-code heartbeat

Troubleshooting Steps

Use the LED state together with the system symptom. If the LED indicates an error, inhibit or temperature condition, perform the corresponding subsystem troubleshooting before replacing the board.

Parts Areas to Check

Beetle Interface Board

For Related Models

Definium AMX 700

Error Number

Cannot obtain worklist

Error Code System Message

​

AMX 700 worklist not working, MWL C-Echo, AE Title port
Question

Why can’t the Definium AMX 700 retrieve a Modality Worklist?

Short Answer

First verify physical/logical networking, then confirm the system and MWL server IP address, AE Title and DICOM port and run C-Echo.

Possible Causes

Network issue; wrong MWL IP; AE Title mismatch/case error; wrong DICOM port; query configuration issue.

Troubleshooting Steps

Document the system AE Title, DICOM port and IP. Open Network Connections, select the MWL server, verify its IP/AE Title/port, then run C-Echo. If C-Echo passes, review the error log and MWL query configuration.

Parts Areas to Check

AMX 700 DICOM configuration; MWL server; hospital network

For Related Models

Definium AMX 700

bottom of page