AMX Troubleshooting & Technical Support
Results Count
Error Number
FPGA_SAFE LED
Error Code System Message
FPGA_SAFE - DS5 YELLOW
Question
What does the FPGA_SAFE LED mean on the GE Definium AMX 700?
Short Answer
Safe-mode FPGA configuration is active.
Possible Causes
Safe-mode FPGA configuration loaded
Troubleshooting Steps
Use the LED state together with the system symptom. If the LED indicates an error, inhibit or temperature condition, perform the corresponding subsystem troubleshooting before replacing the board.
Parts Areas to Check
Beetle Interface Board
For Related Models
Definium AMX 700
Error Number
Fast blinking GREEN
Error Code System Message
Fast blinking GREEN - Image transferring
Question
What does “Fast blinking GREEN” mean on a GE Definium AMX 700?
Short Answer
Fast blinking GREEN: Image transferring.
Possible Causes
Subsystem/category: Digital Detector. Source classification: LED STATUS.
Troubleshooting Steps
Allow transfer to complete; do not move the detector.
Parts Areas to Check
TRAD digital detector
For Related Models
Definium AMX 700
Error Number
Frequency >85 Hz - Video frequency outside monitor specification
Error Code System Message
Frequency >85 Hz - Video frequency outside monitor specification
Question
What does “Frequency >85 Hz” mean on a GE Definium AMX 700?
Short Answer
Frequency >85 Hz: Video frequency outside monitor specification.
Possible Causes
Subsystem/category: Touch Screen Monitor. Source classification: SYSTEM MESSAGE.
Troubleshooting Steps
Restore the approved PC/display configuration and verify the output frequency.
Parts Areas to Check
Monitor; PC display card
For Related Models
Definium AMX 700
Error Number
Frequency out of range
Error Code System Message
Frequency out of monitor specification (>85 Hz)
Question
Why does the Definium AMX 700 monitor report a signal-frequency problem?
Short Answer
The manual states the monitor displays an out-of-range message when the signal frequency exceeds its specification, identified as over 85 Hz.
Possible Causes
Video signal frequency above monitor specification.
Troubleshooting Steps
Verify the PC/display settings and return the video output to the system’s approved configuration before further hardware troubleshooting.
Parts Areas to Check
Mobile PC; display card; monitor
For Related Models
Definium AMX 700
Error Number
Fuse 2A blown
Error Code System Message
2A fuse
Question
How do I identify a blown 2A fuse on the Definium AMX 700 Digital Power Supply?
Short Answer
With proper input voltage present, no voltage at debug-header connection 6 indicates the 2A fuse is the likely cause.
Possible Causes
Blown 2A fuse.
Troubleshooting Steps
Verify proper input voltage is present, then measure debug-header connection 6. If the reading is absent, inspect/replace the 2A fuse using the correct fuse orientation.
Parts Areas to Check
Digital Power Supply; 2A fuse; 12 V converter input branch