Search Results
Search this site
4481 results found with an empty search
- GE 2801127 AMX-4 Plus Exterior Cable Set - 2- 11 | AMCI Medical…
GE 2801127 AMX-4 Plus Exterior Cable Set - 2- 11 for AMX-4, AMX-4+, Definium AMX-700. Available from AMCI Medical Imaging for mobile X-ray service and… GE Healthcare Part No. 2801127 Part Type Special Kit by AMCI Section Area Vertical Column Part Condition New, Harvested. Refurbished Compatible Model AMX-4, AMX-4+, Definium AMX 700 Part Notes Previous Next Request price quote < Search more 2801127 AMX-4 Plus Exterior Cable Set - 2- 11.5 FT" HV Cables, 1- Stator Cable, 1-Main Control Cable. 1 Horizontal Arm, 1 Collimator Cable - COMPLETE SET Part Description
- GE 46-208561P40 Cap screw 3/8-16 X 0 | AMCI Medical Imaging
GE 46-208561P40 Cap screw 3/8-16 X 0 for AMX-4. Available from AMCI Medical Imaging for mobile X-ray service and replacement. GE Healthcare Part No. 46-208561P40 Part Type Fastener Retainer Section Area Vertical Column Part Condition New Compatible Model AMX-4 Part Notes Previous Next Request price quote < Search more 46-208561P40 Cap screw 3/8-16 X 0.625 Long Hex HD Part Description
- Error 0207 | AMCI Medical Imaging
GE mobile X-ray troubleshooting for Error 0207. Review possible causes, diagnostic steps and service guidance from AMCI Medical Imaging. Troubleshooting AMX Error Number Error 0207 Error Code / System Message Error Question What does error 0207 mean on a GE Optima XR240amx? Short Answer Error 0207: Overcurrent detected on filament. Possible Causes Subsystem/category: Fil Large. Manual recovery/error class: INHIBIT. Troubleshooting Steps Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts. Parts Areas to Check < Search more Previous Next For Related Models Optima XR240amx
- GE 5341549 Beetle Djinn Power Module | AMCI Medical Imaging
GE 5341549 Beetle Djinn Power Module for Brivo XR285amx, Optima XR200amx, Optima XR220amx, Optima XR240amx, AMX Navigate. Available from AMCI Medical Imaging… GE Healthcare Part No. 5341549 Part Type Assembly Section Area Thorax Mid Section Part Condition Harvested, Refurbished Compatible Model Brivo XR285amx, Optima XR200amx, Optima XR220amx, Optima XR240amx, AMX Navigate Part Notes Optional Exchange Required Previous Next Request price quote < Search more 5341549 Beetle Djinn Power Module Part Description
- GE 46-270275G1 Drive Handle Assembly | AMCI Medical Imaging
GE 46-270275G1 Drive Handle Assembly for AMX-4. Available from AMCI Medical Imaging for mobile X-ray service and replacement. GE Healthcare Part No. 46-270275G1 Part Type Assembly Section Area Drive Handle Part Condition Harvested, Refurbished Compatible Model AMX-4 Part Notes Optional Exchange Required Previous Next Request price quote < Search more 46-270275G1 Drive Handle Assembly Part Description
- GE 511A590P272 Terminal Female #22 Wire On Wires | AMCI Medical…
GE 511A590P272 Terminal Female #22 Wire On Wires for AMX-4. Available from AMCI Medical Imaging for mobile X-ray service and replacement. GE Healthcare Part No. 511A590P272 Part Type Connector Terminal Section Area Top Cover Part Condition New Compatible Model AMX-4 Part Notes Previous Next Request price quote < Search more 511A590P272 Terminal Female #22 Wire On Wires Part Description
- Error 0x0466 | AMCI Medical Imaging
GE mobile X-ray troubleshooting for Error 0x0466. Review possible causes, diagnostic steps and service guidance from AMCI Medical Imaging. Troubleshooting AMX Error Number Error 0x0466 Error Code / System Message Error Question What does 0x0466 mean on a GE Brivo XR285amx? Short Answer 0x0466: 48V LVPS out of specification. Possible Causes The Djinn Beetle generator reported a Rotor FW fault: 48V LVPS out of specification. Troubleshooting Steps Run Rotation Test and Tube Rotor Impedance. Check rotor communication, tube rotation cable, stator resistance/isolation, Auxiliary LEDs/Neons, 48 V supply, and the error-data value for 0x0480 when applicable. Parts Areas to Check < Search more Previous Next For Related Models Brivo XR285amx
- GE 46-250207P37 Adjust Knob | AMCI Medical Imaging
GE 46-250207P37 Adjust Knob for AMX-4, AMX-4+, Definium AMX 700. Available from AMCI Medical Imaging for mobile X-ray service and replacement. GE Healthcare Part No. 46-250207P37 Part Type Metal Section Area Collimator Part Condition New, Harvested Compatible Model AMX-4, AMX-4+, Definium AMX 700 Part Notes Previous Next Request price quote < Search more 46-250207P37 Adjust Knob Part Description
- GE 2801280 Guard Arm for Collimator - China | AMCI Medical…
GE 2801280 Guard Arm for Collimator - China for AMX-4. Available from AMCI Medical Imaging for mobile X-ray service and replacement. GE Healthcare Part No. 2801280 Part Type Special Kit by AMCI Section Area Collimator Part Condition New, Harvested. Refurbished Compatible Model AMX-4 Part Notes Previous Next Request price quote < Search more 2801280 Guard Arm for Collimator - China Part Description
- Error 45E | AMCI Medical Imaging
GE mobile X-ray troubleshooting for Error 45E. Review possible causes, diagnostic steps and service guidance from AMCI Medical Imaging. Troubleshooting AMX Error Number Error 45E Error Code / System Message ERROR 45E - Filament-short detect asserted after inverter start Error Question What does ERROR 45E mean on a GE AMX-Series mobile X-ray? Short Answer ERROR 45E: Filament-short detect asserted after inverter start. Possible Causes Filament short-detect status did not clear as expected. Troubleshooting Steps Check filament short-detect signal, tube/cable orientation and filament circuit. Parts Areas to Check Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault < Search more Previous Next For Related Models AMX-4, AMX-4+, Definium AMX 700
- GE 5396659 Rotation Lock Base 2 | AMCI Medical Imaging
GE 5396659 Rotation Lock Base 2 for Brivo XR285amx, Optima XR200amx, Optima XR220amx, Optima XR240amx, AMX Navigate. Available from AMCI Medical Imaging for… GE Healthcare Part No. 5396659 Part Type Metal Section Area Collimator Part Condition Harvested Compatible Model Brivo XR285amx, Optima XR200amx, Optima XR220amx, Optima XR240amx, AMX Navigate Part Notes Previous Next Request price quote < Search more 5396659 Rotation Lock Base 2 Part Description
- GE 5350002-4 Spyder System Controller | AMCI Medical Imaging
GE 5350002-4 Spyder System Controller for Optima XR220amx. Available from AMCI Medical Imaging for mobile X-ray service and replacement. GE Healthcare Part No. 5350002-4 Part Type Circuit Board Section Area Thorax Mid Section Part Condition Harvested, Refurbished Compatible Model Optima XR220amx Part Notes Previous Next Request price quote < Search more 5350002-4 Spyder System Controller Part Description