top of page

Search Results

Search this site

4481 results found with an empty search

  • GE 46-170012P64 Lock Nut 1/4-20 Hex | AMCI Medical Imaging

    GE 46-170012P64 Lock Nut 1/4-20 Hex for AMX-4, AMX-4+, Definium AMX 700. Available from AMCI Medical Imaging for mobile X-ray service and replacement. GE Healthcare Part No. 46-170012P64 Part Type Fastener Retainer Section Area Generator Mid Section Part Condition New Compatible Model AMX-4, AMX-4+, Definium AMX 700 Part Notes Previous Next Request price quote < Search more 46-170012P64 Lock Nut 1/4-20 Hex Part Description

  • KVP CAL ERROR 4 | AMCI Medical Imaging

    GE mobile X-ray troubleshooting for KVP CAL ERROR 4. Review possible causes, diagnostic steps and service guidance from AMCI Medical Imaging. Troubleshooting AMX Error Number KVP CAL ERROR 4 Error Code / System Message KVP CAL ERROR 4 - kVp calibration parameter out of limits Error Question What does KVP CAL ERROR 4 mean on a GE AMX-Series mobile X-ray? Short Answer KVP CAL ERROR 4: kVp calibration parameter out of limits. Possible Causes One or more kVp calibration parameters exceeded allowed limits. Troubleshooting Steps Verify kVp values/meter setup and repeat kVp calibration. Parts Areas to Check CPU board; calibration data; subsystem identified by the error < Search more Previous Next For Related Models AMX-4, AMX-4+, Definium AMX 700

  • GE A7000DX 10 x 12 and 14 x 17 Table Battery Charger and AC…

    GE A7000DX 10 x 12 and 14 x 17 Table Battery Charger and AC Cable for AMX Navigate. Available from AMCI Medical Imaging for mobile X-ray service and… GE Healthcare Part No. A7000DX Part Type Detector Related Section Area Storage Bin Part Condition Harvested Compatible Model AMX Navigate Part Notes Previous Next Request price quote < Search more A7000DX 10 x 12 and 14 x 17 Table Battery Charger and AC Cable Part Description

  • GE 46-250662P3 Strap | AMCI Medical Imaging

    GE 46-250662P3 Strap for AMX-4, AMX-4+, Definium AMX 700. Available from AMCI Medical Imaging for mobile X-ray service and replacement. GE Healthcare Part No. 46-250662P3 Part Type Metal Section Area Generator Mid Section Part Condition New, Harvested Compatible Model AMX-4, AMX-4+, Definium AMX 700 Part Notes Previous Next Request price quote < Search more 46-250662P3 Strap Part Description

  • GE 3000-M3C-04 Hexagon Nut M3-0 | AMCI Medical Imaging

    GE 3000-M3C-04 Hexagon Nut M3-0 for Brivo XR285amx, Optima XR200amx, Optima XR220amx, Optima XR240amx, AMX Navigate. Available from AMCI Medical Imaging for… GE Healthcare Part No. 3000-M3C-04 Part Type Fastener Retainer Section Area Multi Part Condition New Compatible Model Brivo XR285amx, Optima XR200amx, Optima XR220amx, Optima XR240amx, AMX Navigate Part Notes Previous Next Request price quote < Search more 3000-M3C-04 Hexagon Nut M3-0.5 Grade 8 Steel Zinc Plated Part Description

  • GE 2000-M5-02 Flat Washer M5 Steel 140HV Zinc Plated | AMCI…

    GE 2000-M5-02 Flat Washer M5 Steel 140HV Zinc Plated for Brivo XR285amx, Optima XR200amx, Optima XR220amx, Optima XR240amx, AMX Navigate. Available from AMCI… GE Healthcare Part No. 2000-M5-02 Part Type Fastener Retainer Common Section Area Multi Part Condition New Compatible Model Brivo XR285amx, Optima XR200amx, Optima XR220amx, Optima XR240amx, AMX Navigate Part Notes Previous Next Request price quote < Search more 2000-M5-02 Flat Washer M5 Steel 140HV Zinc Plated Part Description

  • Tube rotor issues | AMCI Medical Imaging

    GE mobile X-ray troubleshooting for Tube rotor issues. Review possible causes, diagnostic steps and service guidance from AMCI Medical Imaging. Troubleshooting AMX Error Number Tube rotor issues Error Code / System Message Error Question How do I test tube rotor impedance on the Brivo XR285amx? Short Answer The Tube Rotor Impedance diagnostic measures the tube stator winding impedance. Possible Causes Open/shorted stator winding; tube rotation cable issue; rotor-board fault. Troubleshooting Steps Run Tube Rotor Impedance. If abnormal, inspect the rotation cable, then measure stator Common-to-Phase and Common-to-Auxiliary resistance and isolation at the tube side. Parts Areas to Check < Search more Previous Next For Related Models Brivo XR285amx

  • GE 5661227-3 Wireless USB Host Radio Board (UWB Dongle, Anatel…

    GE 5661227-3 Wireless USB Host Radio Board (UWB Dongle, Anatel Label, US only) Rev for Optima XR200amx, Optima XR220amx. Available from AMCI Medical Imaging… GE Healthcare Part No. 5661227-3 Part Type Cable Assembly Section Area Thorax Mid Section Part Condition New, Harvested. Refurbished Compatible Model Optima XR200amx, Optima XR220amx Part Notes Previous Next Request price quote < Search more 5661227-3 Wireless USB Host Radio Board (UWB Dongle, Anatel Label, US only) Rev. 3 Part Description

  • GE 2801261 NEW Type Pulley Cable Bar Assembly for Horizontal…

    GE 2801261 NEW Type Pulley Cable Bar Assembly for Horizontal Arm for AMX-4. Available from AMCI Medical Imaging for mobile X-ray service and replacement. GE Healthcare Part No. 2801261 Part Type Special Kit by AMCI Section Area Vertical Column Part Condition New, Harvested. Refurbished Compatible Model AMX-4 Part Notes Previous Next Request price quote < Search more 2801261 NEW Type Pulley Cable Bar Assembly for Horizontal Arm Part Description

  • GE 46-279290P1 Support Angle 4 Slots | AMCI Medical Imaging

    GE 46-279290P1 Support Angle 4 Slots for AMX-4. Available from AMCI Medical Imaging for mobile X-ray service and replacement. GE Healthcare Part No. 46-279290P1 Part Type Metal Section Area Base Part Condition New, Harvested Compatible Model AMX-4 Part Notes Previous Next Request price quote < Search more 46-279290P1 Support Angle 4 Slots Part Description

  • GE 46-170015P19 Screw Lock 6-32 X 0 | AMCI Medical Imaging

    GE 46-170015P19 Screw Lock 6-32 X 0 for AMX-4, AMX-4+, Definium AMX 700. Available from AMCI Medical Imaging for mobile X-ray service and replacement. GE Healthcare Part No. 46-170015P19 Part Type Fastener Retainer Common Section Area Generator Mid Section Part Condition New Compatible Model AMX-4, AMX-4+, Definium AMX 700 Part Notes Previous Next Request price quote < Search more 46-170015P19 Screw Lock 6-32 X 0.38 Long Bind HD Part Description

  • GE 2243998 CPU PWB with Software 3 | AMCI Medical Imaging

    GE 2243998 CPU PWB with Software 3 for AMX-4. Available from AMCI Medical Imaging for mobile X-ray service and replacement. GE Healthcare Part No. 2243998 Part Type Application Firmware Software Section Area Generator Mid Section Part Condition New, Harvested. Refurbished Compatible Model AMX-4 Part Notes Previous Next Request price quote < Search more 2243998 CPU PWB with Software 3.0 Part Description