top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

After stored now behaves abnormally

Error Code System Message

​

XR220 stored system low battery, XR200 cold storage errors
Question

What should I do if an XR200amx or XR220amx has been stored and now behaves abnormally?

Short Answer

A system stored at low temperature or with batteries connected/discharged can show low-temperature, low-charge, shutdown and generator errors without necessarily requiring parts.

Possible Causes

Cold system; self-discharge; very low state of charge; improper storage.

Troubleshooting Steps

Allow a cold system to stabilize to the specified environment. Fully recharge it using the approved charging practice, then reassess errors before replacing batteries or generator components.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

BDM mismatch screen

Error Code System Message

​

XR200 BDM mismatch, XR220 restore cancel, system self recovery
Question

What should I do when an XR200amx or XR220amx displays a BDM Mismatch screen?

Short Answer

On a normally operating system where no parts were changed, the Exclusive manual instructs the technician to select RESTORE.

Possible Causes

Board Data Module fingerprint or firmware mismatch detected at power-up.

Troubleshooting Steps

First confirm whether any intelligent board was replaced. If no parts were changed, select RESTORE. If a board was replaced, follow the replacement finalization instructions for the correct Restore/Cancel choice.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Battery pack needs to be replace

Error Code System Message

​

XR220 battery integration, XR200 replacement pack, safe charge mode
Question

How should a replacement battery pack be integrated into XR200amx/XR220amx?

Short Answer

The Exclusive manual recommends a two-step process so packs of different state-of-charge or vintage can equalize safely.

Possible Causes

SOC mismatch between new and existing packs.

Troubleshooting Steps

Use the approved Safe Charge mode for an extended undisturbed charge, then Normal Charge mode until one hour after the battery meter reaches 100%. Follow the replacement procedure for exact jumper/service steps.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Black screen

Error Code System Message

​

XR220 black screen, XR200 no video, PC Spyder cable
Question

What should I check for a black screen on XR200amx or XR220amx?

Short Answer

Use PC Boot View and an external VGA monitor to separate a PC video failure from a Spyder/video-cable path problem.

Possible Causes

PC failure; PC-to-Spyder video cable fault; Spyder video path fault.

Troubleshooting Steps

Enter the approved degraded-mode workflow, check PC power, connect an external monitor, and reseat the PC-to-Spyder video cable. Replace hardware only after isolating the failing path.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Boot stop after a board was replaced

Error Code System Message

​

XR220 boot halted after board replacement, BDM recovery, Load from Cold
Question

Why did my XR200amx or XR220amx boot stop after a board replacement?

Short Answer

The BDM checks component fingerprint and firmware data at power-up; a mismatch can halt boot until the correct recovery action is selected.

Possible Causes

New/replaced intelligent board does not match BDM fingerprint or firmware data.

Troubleshooting Steps

Identify the replaced part and use the manual's Restore/Cancel table for that component. If an incorrect choice was previously made, a Load from Cold may be required.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx