AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 426
Error Code System Message
ERROR 426 - Autocalibration data fault
Question
What does ERROR 426 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 426: Autocalibration data fault.
Possible Causes
Auto-calibration data indicated a calibration error or potentially invalid information.
Troubleshooting Steps
Review calibration status and repeat the required generator calibration.
Parts Areas to Check
Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
Error 42F
Error Code System Message
42F
Question
What should I check for error 42f on the AMX-4 Series?
Short Answer
1) Defective RAM on the CPU Board 2) Defective CPU Board, Bad Rotor Controller Board
Possible Causes
1) Defective RAM on the CPU Board 2) Defective CPU Board 3) Bad Rotor Controller Board
Troubleshooting Steps
1) Defective RAM on the CPU Board 2) Defective CPU Board. 3) Check for possible bad Rotor Controller Board
Parts Areas to Check
CPU Board
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
Error 430
Error Code System Message
ERROR 430 - Rotor frequency incorrect during prep
Question
What does ERROR 430 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 430: Rotor frequency incorrect during prep.
Possible Causes
Rotor/inverter feedback frequency was incorrect.
Troubleshooting Steps
Check rotor-inverter feedback, generator controller, CPU connection and inverter circuit.
Parts Areas to Check
Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
Error 431
Error Code System Message
ERROR 431 - kVp demand feedback too high
Question
What does ERROR 431 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 431: kVp demand feedback too high.
Possible Causes
kVp demand feedback exceeded expected command during prep.
Troubleshooting Steps
Check CPU-generator controller connection and kVp demand/control path.
Parts Areas to Check
Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
Error 432
Error Code System Message
ERROR 432 - Filament-current demand feedback too high
Question
What does ERROR 432 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 432: Filament-current demand feedback too high.
Possible Causes
Filament current demand feedback exceeded expected value during prep.
Troubleshooting Steps
Check filament board/CPU connection, current feedback and tube filament connections.
Parts Areas to Check
Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault