top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 459

Error Code System Message

459

AMX-4, AMX-4+, Definium AMX 700, ERROR 459, Error 459
Question

What should I check for error 459 on the AMX-4 Series?

Short Answer

1. Check the Cable Connection From The CPU Board To The Filament kVp Board – Bad Ribbon Cable (Orderable part # 46-302366P2)

Possible Causes

1. Check the Cable Connection From The CPU Board To The Filament kVp Board – Bad Ribbon Cable (Orderable part # 46-302366P2)

Troubleshooting Steps

1. Check the Cable Connection From The CPU Board To The Filament kVp Board – Bad Ribbon Cable (Orderable part # 46-302366P2)

Parts Areas to Check

CPU Board; Filament kVp Board

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 45A

Error Code System Message

ERROR 45A - Leakage-compensation feedback low

ERROR 45A, Leakage-compensation feedback low, AMX-4 ERROR 45A, AMX-4+ ERROR 45A, Definium AMX 700 ERROR 45A, GE AMX error ERROR 45A
Question

What does ERROR 45A mean on a GE AMX-Series mobile X-ray?

Short Answer

ERROR 45A: Leakage-compensation feedback low.

Possible Causes

Leakage-compensation feedback was lower than expected.

Troubleshooting Steps

Check leakage compensation feedback and filament-board/kVp feedback path.

Parts Areas to Check

Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 45A

Error Code System Message

45A

AMX-4, AMX-4+, Definium AMX 700, ERROR 45A, Error 45A
Question

What should I check for error 45A on the AMX-4 Series?

Short Answer

1. Defective Filament kVp Control Board

Possible Causes

1. Defective Filament kVp Control Board

Troubleshooting Steps

1. Defective Filament kVp Control Board

Parts Areas to Check

​

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 45B

Error Code System Message

45B

AMX-4, AMX-4+, Definium AMX 700, ERROR 45B, Error 45B
Question

What should I check for error 45B on the AMX-4 Series?

Short Answer

1. Filament kVp Board may be defective

Possible Causes

1. Filament kVp Board may be defective

Troubleshooting Steps

1. Filament kVp Board may be defective

Parts Areas to Check

Filament kVp Board

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 45B

Error Code System Message

ERROR 45B - Filament feedback low

ERROR 45B, Filament feedback low, AMX-4 ERROR 45B, AMX-4+ ERROR 45B, Definium AMX 700 ERROR 45B, GE AMX error ERROR 45B
Question

What does ERROR 45B mean on a GE AMX-Series mobile X-ray?

Short Answer

ERROR 45B: Filament feedback low.

Possible Causes

Filament feedback was lower than command.

Troubleshooting Steps

Check filament feedback, tube filament/cathode connections and filament board.

Parts Areas to Check

Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault

For Related Models

AMX-4, AMX-4+, Definium AMX 700

bottom of page