AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 459
Error Code System Message
459
Question
What should I check for error 459 on the AMX-4 Series?
Short Answer
1. Check the Cable Connection From The CPU Board To The Filament kVp Board – Bad Ribbon Cable (Orderable part # 46-302366P2)
Possible Causes
1. Check the Cable Connection From The CPU Board To The Filament kVp Board – Bad Ribbon Cable (Orderable part # 46-302366P2)
Troubleshooting Steps
1. Check the Cable Connection From The CPU Board To The Filament kVp Board – Bad Ribbon Cable (Orderable part # 46-302366P2)
Parts Areas to Check
CPU Board; Filament kVp Board
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
Error 45A
Error Code System Message
ERROR 45A - Leakage-compensation feedback low
Question
What does ERROR 45A mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 45A: Leakage-compensation feedback low.
Possible Causes
Leakage-compensation feedback was lower than expected.
Troubleshooting Steps
Check leakage compensation feedback and filament-board/kVp feedback path.
Parts Areas to Check
Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
Error 45A
Error Code System Message
45A
Question
What should I check for error 45A on the AMX-4 Series?
Short Answer
1. Defective Filament kVp Control Board
Possible Causes
1. Defective Filament kVp Control Board
Troubleshooting Steps
1. Defective Filament kVp Control Board
Parts Areas to Check
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
Error 45B
Error Code System Message
45B
Question
What should I check for error 45B on the AMX-4 Series?
Short Answer
1. Filament kVp Board may be defective
Possible Causes
1. Filament kVp Board may be defective
Troubleshooting Steps
1. Filament kVp Board may be defective
Parts Areas to Check
Filament kVp Board
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
Error 45B
Error Code System Message
ERROR 45B - Filament feedback low
Question
What does ERROR 45B mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 45B: Filament feedback low.
Possible Causes
Filament feedback was lower than command.
Troubleshooting Steps
Check filament feedback, tube filament/cathode connections and filament board.
Parts Areas to Check
Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault