AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 45E

Error Code System Message

ERROR 45E - Filament-short detect asserted after inverter start

ERROR 45E, Filament-short detect asserted after inverter start, AMX-4 ERROR 45E, AMX-4+ ERROR 45E, Definium AMX 700 ERROR 45E, GE AMX error ERROR 45E
Question

What does ERROR 45E mean on a GE AMX-Series mobile X-ray?

Short Answer

ERROR 45E: Filament-short detect asserted after inverter start.

Possible Causes

Filament short-detect status did not clear as expected.

Troubleshooting Steps

Check filament short-detect signal, tube/cable orientation and filament circuit.

Parts Areas to Check

Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 45F

Error Code System Message

45F

AMX-4, AMX-4+, Definium AMX 700, AMCI-AMX4-TS-0175, Error code 45F. Apply to recent Filament kVp Board part # 46-264986G2, the electronic components on the board capacitor C98, C233 and C234 failing on the board due to non-qualified vendor (Electrocube) which are suspecte
Question

How do I troubleshoot this AMX-4 Series issue: Error code 45F. Apply to recent Filament kVp Board part # 46-264986G2, the electronic components on the board capacitor C98, C233 and C234 failing on the board due to non-qualified vendor (Electrocube) which are suspecte?

Short Answer

change out capacitors on Filament kVp Board part # 46-264986G2, order Part 46-170087P100, Capacitor C98 and order Part # 46-170087P4 Capacitor C233 and C234. Refer to Service Note 3999 on replacement instructions.

Possible Causes

27) Problem: Error code 45F. Apply to recent Filament kVp Board part # 46-264986G2, the electronic components on the board capacitor C98, C233 and C234 failing on the board due to non-qualified vendor (Electrocube) which are suspected to have heated-related quality issues.

Troubleshooting Steps

change out capacitors on Filament kVp Board part # 46-264986G2, order Part 46-170087P100, Capacitor C98 and order Part # 46-170087P4 Capacitor C233 and C234. Refer to Service Note 3999 on replacement instructions.

Parts Areas to Check

Filament kVp Board

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 45F

Error Code System Message

45F

AMX-4, AMX-4+, Definium AMX 700, ERROR 45F, Error 45F
Question

What should I check for error 45F on the AMX-4 Series?

Short Answer

1) Check correct wiring between J5 plug on Filament kVp Board and HV Transformer Tank. 2) Check spark gaps on HV Transformer Tank (0.010 in.). 3) Check that HV cables are not reversed. 4) Check wiring of J4 plug to Filament kVp Board. 5) Check that U18 on Filament kVp Board is #1489, not #14C89. 6) Check/replace Filament kVp Board. 7) Check FIL SHORTED at J2-25 on Filament kVp Board. 8) Check for insufficient filament drive: XS-1 to XS-2 at HV Transformer Tank should be approximately 17 VAC at 52 kVp with PREP. 9) Check for loose J3 on Filter Board.

Possible Causes

Apply to recent Filament kVp Board part # 46-264986G2, the electronic components on the board capacitor C98, C233 and C234 failing on the board due to non-qualified vendor (Electrocube) which are suspected to have heated-related quality issues. Refer to Service Note 3999 1. Check for Correct Wiring Between J5 Plug On Filament kVp Board And HV Transformer Tank 2. Check Spark Gaps On HV Transformer Tank (0.010”) 3. Check that HV Cables Are Not Reversed 4. Check Wiring Of J4 Plug To Filament kVp Board 5. Check that U18 On Filament kVp Board Is #1489 — NOT #14C89 6. Defective Filament kVp Control Board (See Additional Notes #3) 7. FIL SHORTED (Check J2-25 On Filament kVp Board) 8. Insufficient Filament Drive (XS-1 To XS-2 [HV Transformer Tank] = Approximately 17 VAC @ 52 kVp With PREP) 9. Loose J3 On Filter Board

Troubleshooting Steps

Apply to recent Filament kVp Board part # 46-264986G2, the electronic components on the board capacitor C98, C233 and C234 failing on the board due to non-qualified vendor (Electrocube) which are suspected to have heated-related quality issues. Refer to Service Note 3999 1. Check for Correct Wiring Between J5 Plug On Filament kVp Board And HV Transformer Tank 2. Check Spark Gaps On HV Transformer Tank (0.010”) 3. Check that HV Cables Are Not Reversed 4. Check Wiring Of J4 Plug To Filament kVp Board 5. Check that U18 On Filament kVp Board Is #1489 — NOT #14C89 6. Defective Filament kVp Control Board (See Additional Notes #3) 7. FIL SHORTED (Check J2-25 On Filament kVp Board) 8. Insufficient Filament Drive (XS-1 To XS-2 [HV Transformer Tank] = Approximately 17 VAC @ 52 kVp With PREP) 9. Loose J3 On Filter Board

Parts Areas to Check

Filament kVp Board; HV Transformer Tank

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 45F

Error Code System Message

ERROR 45F - Filament-short signal remains active after filaments removed

ERROR 45F, Filament-short signal remains active after filaments removed, AMX-4 ERROR 45F, AMX-4+ ERROR 45F, Definium AMX 700 ERROR 45F, GE AMX error ERROR 45F
Question

What does ERROR 45F mean on a GE AMX-Series mobile X-ray?

Short Answer

ERROR 45F: Filament-short signal remains active after filaments removed.

Possible Causes

Short-detect remained active after filament loads were removed.

Troubleshooting Steps

Inspect high-voltage cable polarity/orientation and filament short-detect circuitry.

Parts Areas to Check

Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 460

Error Code System Message

ERROR 460 - Tube pressure switch open during exposure interlock

ERROR 460, Tube pressure switch open during exposure interlock, AMX-4 ERROR 460, AMX-4+ ERROR 460, Definium AMX 700 ERROR 460, GE AMX error ERROR 460
Question

What does ERROR 460 mean on a GE AMX-Series mobile X-ray?

Short Answer

ERROR 460: Tube pressure switch open during exposure interlock.

Possible Causes

Tube-pressure/low-oil status indicated an open condition.

Troubleshooting Steps

Check tube pressure switch and related circuitry; evaluate tube/HV tank condition.

Parts Areas to Check

Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault

For Related Models

AMX-4, AMX-4+, Definium AMX 700