Error 461

Error Code System Message

ERROR 461 - 60 Hz inverter OK signal low during exposure interlock

ERROR 461, 60 Hz inverter OK signal low during exposure interlock, AMX-4 ERROR 461, AMX-4+ ERROR 461, Definium AMX 700 ERROR 461, GE AMX error ERROR 461
Question

What does ERROR 461 mean on a GE AMX-Series mobile X-ray?

Short Answer

ERROR 461: 60 Hz inverter OK signal low during exposure interlock.

Possible Causes

60 Hz inverter OK status was not present.

Troubleshooting Steps

Check inverter OK/status line, CPU input and inverter/relay circuitry.

Parts Areas to Check

Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 462

Error Code System Message

462

AMX-4, AMX-4+, Definium AMX 700, ERROR 462, Error 462
Question

What should I check for error 462 on the AMX-4 Series?

Short Answer

1) Check wires to Power Module SCR Q3 block on AMX1A4 panel for loose connections. 2) Check mA metering circuit/calibration: AR125 Pin 1 or TP-3 (ref. TP-5) on Filament kVp Board = 4.7 V/100 mA. 3) Batteries may be soft; refer to Error 465; could be one bad cell up to full pack. 4) Check High Voltage Bleeder. 5) Check Battery Charger Board or open fuse F4. 6) Errors 462 and 466 with breaker tripping on exposures above 48 kVp: check/replace HV Transformer Tank. 7) Error 462 on exposures >70 kVp with Error 439 closed-loop resistance: recalibrate unit. 8) Error 462 during calibration with no exposure termination, but normal termination in Check-Sum Bypass mode: spark gap may be shorted. 9) Check Filament kVp Board J3 pins 3 and 4 for high-resistance jumper. 10) Check battery connections.

Possible Causes

1. Check Wires To Power Module SCR Q3 Block On AMX1A4 Panel For Loose Connections 2. Faulty MA Metering Circuit - Or Out Of Calibration– (AR125, Pin 1 Or TP-3 [Ref. TP-5] On Filament kVp Board = 4.7V/100MA) 3. Batteries May Be Soft - Refer To ERROR 465 - could be 1 Bad Cell Up To Full Pack 4. could be Faulty High Voltage Bleeder 5. Battery Charger Board Failure Or Fuse F4 Open 6. Errors 462 And 466 - Tripping Breaker With Exposures Above 48 kVp = Replace HV Transformer Tank 7. Error 462 On Exposures >70 kVp And Error 439 Closed Loop Resistance? = Re-Calibrate Unit 8. ERROR 462 During Calibration With NO Exposure Termination (Counts Actual Time)—In Check-Sum Bypass Mode, The Exposure Terminates Normally (Uses Time Selection only). Solution: SPARK GAP SHORTED. 9. Filament kVp Board J3 Pins 3 And 4 (Jumper Has High Resistance) 10. Check for loose battery connections.

Troubleshooting Steps

1. Check Wires To Power Module SCR Q3 Block On AMX1A4 Panel For Loose Connections 2. Faulty MA Metering Circuit - Or Out Of Calibration– (AR125, Pin 1 Or TP-3 [Ref. TP-5] On Filament kVp Board = 4.7V/100MA) 3. Batteries May Be Soft - Refer To ERROR 465 - could be 1 Bad Cell Up To Full Pack 4. could be Faulty High Voltage Bleeder 5. Battery Charger Board Failure Or Fuse F4 Open 6. Errors 462 And 466 - Tripping Breaker With Exposures Above 48 kVp = Replace HV Transformer Tank 7. Error 462 On Exposures >70 kVp And Error 439 Closed Loop Resistance? = Re-Calibrate Unit 8. ERROR 462 During Calibration With NO Exposure Termination (Counts Actual Time)—In Check-Sum Bypass Mode, The Exposure Terminates Normally (Uses Time Selection only). Solution: SPARK GAP SHORTED. 9. Filament kVp Board J3 Pins 3 And 4 (Jumper Has High Resistance) 10. Check for loose battery connections.

Parts Areas to Check

Battery Charger Board; Filament kVp Board; HV Transformer Tank; Power Module SCR Q3; Battery

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 462

Error Code System Message

ERROR 462 - Exposure time too long

ERROR 462, Exposure time too long, AMX-4 ERROR 462, AMX-4+ ERROR 462, Definium AMX 700 ERROR 462, GE AMX error ERROR 462
Question

What does ERROR 462 mean on a GE AMX-Series mobile X-ray?

Short Answer

ERROR 462: Exposure time too long.

Possible Causes

Calculated exposure time exceeded the maximum allowed value.

Troubleshooting Steps

Check selected technique and battery voltage; correct conditions before retrying exposure.

Parts Areas to Check

Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 463

Error Code System Message

ERROR 463 - kVp command feedback low during exposure interlock

ERROR 463, kVp command feedback low during exposure interlock, AMX-4 ERROR 463, AMX-4+ ERROR 463, Definium AMX 700 ERROR 463, GE AMX error ERROR 463
Question

What does ERROR 463 mean on a GE AMX-Series mobile X-ray?

Short Answer

ERROR 463: kVp command feedback low during exposure interlock.

Possible Causes

kVp demand feedback was low during exposure interlock.

Troubleshooting Steps

Check CPU-generator connection, filament board and feedback path.

Parts Areas to Check

Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 463

Error Code System Message

463

AMX-4, AMX-4+, Definium AMX 700, ERROR 463, Error 463
Question

What should I check for error 463 on the AMX-4 Series?

Short Answer

1. Defective CPU Board 2. Defective Filament kVp Control Board 3. Error 463 And 469 - Loose Connection On/Or Bad 1 KHz Inverter Driver Board 4. Safety Contactor Contact Issue

Possible Causes

1. Defective CPU Board 2. Defective Filament kVp Control Board 3. Error 463 And 469 - Loose Connection On/Or Bad 1 KHz Inverter Driver Board 4. Safety Contactor Contact Issue

Troubleshooting Steps

1. Defective CPU Board 2. Defective Filament kVp Control Board 3. Error 463 And 469 - Loose Connection On/Or Bad 1 KHz Inverter Driver Board 4. Safety Contactor Contact Issue

Parts Areas to Check

CPU Board; 1 KHZ Inverter Driver Board; Safety Contactor

For Related Models

AMX-4, AMX-4+, Definium AMX 700